button to select the project would you
want calibration. ( Select the order of
0.00m/s→10.00m/s→33.00%RH→75.00%RH→
25.00 )
A. Wind speed calibration (0.00m/s)
Press
button to select the 0.00m/s and
press
button to st a rt and the 0.00m/s is
blinking.
The duration of calibration time is 5 minutes.
After end of the 0m/s calibration, you can press
button to shut down or press
button to into 10 m/s calibration.
B. Wind speed calibration (10.00m/s)
Put the probe into the standard 10m/s wind
tunnel, and press
button to select the
10.00m/s and press
button to st a rt
calibration 10m/s.
The duration of calibration time is 5 minutes.
After end of the 10m/s calibration, then will be
into the normal display.
C. Humidity calibration (33.0%RH)
Put the probe in the 33.0%rh chamber and
press
button to select the 3 3.0%RH and
press
button to st a rt calibration