Equinox Systems EQ-TM- 4001 User Manual page 4

Hot-wire air velocity meter
Table of Contents

Advertisement

• Meter Dimension
: 156x73x35mm(Lx W x H).
• Probe Dimension :
Wire length : 160cm
Probe length : 128cm
Probe diameter of tip : 1.6cm
Probe diameter of base : 3.2cm
• Meter weight
: 250g.(without batteries)
• Probe Weight
: 240 g.
• Po wer supply
: UM-4(AAA)1.5V*6PCS or
AC100~240V DC9V/0.5A(9mm).
• Battery life
: 10 hours.
• Operating T emperature & Humidity:
-20°C to +50°C, <80 %RH.
• Storage Tempe rature & Humidity :
-10°C to +50°C, <70 %RH.
• Data outpu t
: USB PC serial interface. (TM-4003)
• Display
: 60mm(W) x 47mm(L) monochrome
LCD and backlight.
• Standard Accessories :
UM-4 (AAA)1.5V battery X 6 Pcs
33.0%RH.
The duration of calibration time is 10 minutes.
After end of the 33.0%RH calibration, you can
press
button to shut do wn or press
button to into 75%RH calibration.
D. Humidity calibration (75%RH)
Put the probe in the 75%RH chamber and
press
button to select the 7 5%RH and
press
button to st a rt 75%RH calibration.
The duration of calibration time is 10 minutes.
After end of the 75%RH calibration, then will be
into the normal display.
E. Temperature calibration (25.0 )
1. Press
button to select the 25.0
press
button to st a rt 25.0
2. Confirm with the standard value of the error,
press the
and
button to ad just the
temperature and the display will flash.
3. Press
to compl e te the adjustment
and into the normal display.
19. PRECAUTIONS
• Keep the m eter away from electromagnetic
and
calibration.