User Troubleshooting Table; Do Not Ignore Alarms; Power Off The Device - CAIRE Eclipse 5 User Manual

With autosat
Hide thumbs Also See for Eclipse 5:
Table of Contents

Advertisement

Eclipse 5
to the device.
Depress the push-button on the
Universal Cart handle to adjust
the height of the handle.
Step 8:

Power Off the Device

Press and hold the "ON/OFF" Button for two (2)

User Troubleshooting Table

Do Not Ignore Alarms

WARNING: THE MANUFACTURER RECOMMENDS AN ALTERNATE SOURCE OF SUPPLEMEN-
TAL OXYGEN IN THE EVENT OF A POWER OUTAGE, ALARM CONDITION, OR MECHANICAL
FAILURE. CONSULT YOUR PHYSICIAN OR EQUIPMENT PROVIDER FOR THE TYPE OF RESERVE
SYSTEM REQUIRED.
Symptom
The yellow light is
on solid. There is no
audible alarm.
The yellow light is
on. Battery symbol is
flashing on LCD.
The yellow light is
on. Battery symbol is
flashing on LCD.
The yellow light is on.
The green light is off.
20 - ENG
PN 21182453 C | User Manual
Yellow Light
Possible Cause
Your Action
1) Remove power cartridge and re-install to ensure that it is
The Eclipse 5 has de-
tected a problem with
2) If possible, install a separate power cartridge in the Eclipse
the connection to the
power cartridge.
3) If this problem persists, contact healthcare provider.
1) Re-attach the Eclipse 5 to external power or install a
The Eclipse 5 power
cartridge may be
2) Re-charge the power cartridge using AC power
warm.
3) Replace the power cartridge if another is available.
4) If the condition persists, contact your healthcare provider.
The Eclipse 5 power
1) Plug Eclipse 5 into AC power supply and recharge or replace
cartridge voltage may
be low.
1) Clean and replace cabinet inlet filter
2) Ensure that the cannula is not kinked or blocked. If used
The Eclipse 5 may
have detected a flow
3) Ensure that the Eclipse 5 has proper ventilation. It needs to
restriction causing
low or blocked flow.
4) If the problem persists, switch to an alternate source of
seconds to power OFF your Eclipse 5.
Store the Eclipse 5 in a cool and dry location for a
period no longer than 1 month. The Eclipse 5 must
run for a minimum of 2 hours per month to ensure
proper function of the unit.
Note: DO NOT operate the Eclipse 5 and ambulate
while a humidifier is attached. Remove the humidi-
fier bottle before walking. DO NOT lay the Eclipse 5
down while attached to a humidifier bottle.
secure in the concentrator.
5.
fully-charged power cartridge in the Eclipse 5. Allow warm
power cartridge to cool outside the concentrator for 30
minutes.
power cartridge with fully charged power cartridge.
with a humidifier bottle, ensure that it is filled properly and
not creating a blockage.
be at least 3 inches from any surface to ensure the vents are
not blocked.
oxygen and contact healthcare provider for assistance.