CALIBRATION
9.0
Remove all weight from the pan before beginning.
• When in the normal weighing mode, hold [Zero] and press the [Tare] key to
enter the parameter menu as shown in section 9.0..
• The display will show the revision number in the "Weight" display ,The "Unit
Weight" display will show "- - - -
entered.
• The default password is "0000" but other numbers can be set using the
parameter menus. Press the [0] key four times. The display will jump to the
first parameter. "FILt ?" with the Unit weight window showing the current
filter setting.
• Press [M+/↵ ↵ ↵ ↵ ] 8 times until the display shows " " " " CALI". Note that the display will
also show " " " " CALI" "Unit ?" " xx" for selecting the unit of weight to be used
when calibrating after 6 presses.
change the weighing unit during calibration.
• Press the numeric keys to enter the calibrate mass value. For kilograms and
pounds enter the value as a number, for example "3" for 3 kilograms. For
grams enter the weight in grams, i.e. "1000" grams.
• Place the calibration mass on the weighing platform.
The suggested calibrate values for all model :
2kg or 3kg
5lb or 6lb
However any whole value of weight can be used by entering the correct value.
• Press the [M+/↵ ↵ ↵ ↵ ] key,
• The display will show "PASS",
• Press the [M+/↵ ↵ ↵ ↵ ] key to return to normal weighing.
• If the calibration is outside of the previous calibration value by more than
5% ,it will show "FAIL L" or "FAIL H" indicating the new calibration weight
was too low(FAIL L) or too high (FAIL H) in respect to the previous
calibration value.
© Inscale Ltd. 2011
LBC 3
LBC 6
5kg or 6kg
10lb or 12lb
p a g e
" requesting the password number to be
Ignore this function unless yu want to
LBC 15
10kg or 15kg
20lb or 25lb
14
LBC 30
20kg or 30kg
40lb or 50lb