4. Other Operations
4.10 Initializing Settings
If the TS-3200A hangs up because of defective settings, it is necessary to initialize the backup memory
with the following procedure.
1 Initializing the entire backup memory
Turn the power ON while holding down the [TRIG] and [ESC] keys.
2 Initializing settings other than the analog calibration data
Turn the power ON while holding down the [ENTER] and [LOCK] keys.
3 Initializing only the selected condition No.
Turn the power ON while holding down the [TRIG] and [ENTER] keys.
4-21