A
B
Confirm CAL C--LIMIT
Turn Cal Station On
If the LCD display reads CAL C--LIMIT,
1.
Press the "POWER" button on the cal
press the Power/Mode button on the gas
station.
monitor to continue to TRANSMIT.
2.
Seat the GX-2009 into the cal station.
If CAL C--LIMIT is not displayed continue to
the next step.
Then turn on the GX-2009 by pressing
the "POWER/MODE" button on the gas
monitor.
A
B
Calibrate
Calibration Gas Readings
Press the "CAL" button to start calibration.
During calibration, the display shows
current gas readings. Typical calibration
Note: Ensure calibration gas cylinder is connected
will take 3 minutes.
to a demand flow regulator and then to the back
of the cal station. Make sure calibration gas
concentrations match values shown on display.
A
C
Instrument Connected
Once on, the display will show gas channels
Press the "Bump" button to start bump test.
and a transmit message. The gas monitor is
Note: Ensure calibration gas cylinder is connected
ready for a Bump Test or Calibration.
to a demand flow regulator and then to the back
of the cal station. Make sure calibration gas
concentrations match values shown on display.
C
Calibration Results
Once calibration is complete, the display either shows a "P" for pass or an "F" for fail for each
sensor. The display will alternate between the gas concentration and the letter "P" when the
monitor passes calibration. The "CAL" LED lights in green when calibration is passed. "CAL"
LED lights in red if there is a failure.
RKI Instruments, Inc.
•
33248 Central Ave. Union City, CA 94587
www.rkiinstruments.com
B
Bump Test
Bump Gas Reading
During Bump test, the display
shows current gas reading.
D
E
Removing from SDM-2009U
After a successful bump test or calibration,
the GX-2009 will automatically turn off within
15 seconds.
the SDM-2009 "Edit Enter" and "Power"
buttons simultaneously.
Warning: Do not remove the GX-2009
before turning it off, and do not use the
GX-2009's power button to power it down.
•
Phone (800) 754-5165
•
(510) 441-5656
•
Fax (510) 441-5650
C
D
Bump Results
Return to Transmit Screen
The BUMP LED will stop blinking and be on
Press the "EDIT/ENTER" button.
steadily green if the bump test passed or
See step "E" below for turning off
steadily red if the bump test failed. Display
instrument after completion of
will show "P" for pass or "F" for fail for each
bump test.
sensor. The display will alternate between
the gas concentration and the letter "P"
when the monitor passes bump test.
A
Data Transfer To USB Flash Drive
With USB flash drive installed and unit powered on,
press and release the COPY button. The COPY LED will
To turn off manually, press
become solid red while the records in the calibration
station's memory are copied to the flash drive. In
addition, the flash drive's LED will begin to blink.
When the LED stops blinking, the copy operation is
complete. Remove the flash drive from the USB port.
Transfer data to PC with Calibration Station software,
following the instructions in the Operator's Manual.