Ambient temperature range
!
Storage temperature
Ambient class
Degree of protection
Shock resistance
Vibration resistance
CIP cleaning
SIP cleaning
Electromagnetic
compatibility (EMC)
Medium temperature range
Medium density
Medium pressure range
(nominal pressure)
20
Environment
Sensor, transmitter
• Standard: –20 to +60 °C (–4 to +140 °F)
• Optional: –40 to +60 °C (–40 to +140 °F)
Note!
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired.
–40 to +80 °C (–40 to +175 °F), preferably +20 °C (+68 °F)
B, C, I
Standard: IP 67 (NEMA 4X) for transmitter and sensor
According to IEC 68-2-31
Acceleration up to 1 g, 10 to 150 Hz, following IEC 68-2-6
yes
yes
To ICE/EN 61326 and NAMUR recommendation NE 21
Process
Sensor
• –50 to +200 °C (–58 to +392 °F)
• High-temperature version: –50 to +350 °C (–58 to +660 °F)
0 to 5000 kg/m
(0 to 312 lb/ft
Flanges
• Standard version:
– according to DIN PN 16 to 100
– according to ASME B16.5 Cl 150, Cl 300, Cl 600
– JIS 10K, 20K, 40K, 63K
• High-temperature version:
– according to DIN PN 40, 64, 100
– according to ASME B16.5 Cl 150, Cl 300, Cl 600
– JIS 10K, 20K, 63K
)
Proline Promass 84F
Endress+Hauser