Description Safety The Greenlee CLM-1000 and CLM-1000E Cable Length Safety is essential in the use and maintenance of Greenlee Meters are hand-held testing devices with the follow- tools and equipment. This instruction manual and any ing measurement capabilities: resistance, temperature, markings on the tool provide information for avoiding and length.
CLM-1000 • CLM-1000E Important Safety Information SAFETY ALERT Electric shock and fire hazard: SYMBOL • Do not connect this unit to live voltage. • Do not expose this unit to rain or moisture. This symbol is used to call your attention to hazards •...
Page 4
Identification 1. Input terminals, two per Kelvin clips for a total of four Display Icons terminals 1. Low battery indicator 2. LCD 2. User-selected switch position 3. Refer to explanation under “Using the Features.” 3. Ambient temperature indication 4. Rotary switch 4.
CLM-1000 • CLM-1000E Using the Features • HOLD: Press momentarily to enter data hold mode. • : Press momentarily to turn the backlight ON. Once When button is pressed, “dH” is present on the LCD and lit, press button again to turn the backlight OFF. The the current value is captured.
Operation 5. To measure odd cable sizes: • AWG (5, 7, 9, 11 AWG, etc.) (CLM-1000): When measuring odd numbered cable sizes, set the rotary Electric shock hazard: switch to the next smaller size, and multiply the Do not connect the unit to live voltage.
CLM-1000 • CLM-1000E Typical Measurements Typical Configuration Note: Make sure that the clips are as close together as possible in order to improve calibration accuracy. Calibration Bar Calibration 1.800.561.8187 information@itm.com www. .com...
Page 8
Accuracy Length in feet Gauges Diameter (mm Accuracy Length Range Accuracy (CLM-1000) (CLM-1000E) Range 15 ft–500.0 ft ± (1.0% + 5.0 ft) 1000.0 ft 500.1 ft–999.9 ft ± (1.0% + 9.0 ft) 18 AWG–2 AWG 0.75–35 10,000 ft 1000 ft–9999 ft ±...
APO Timing: Idle for 10 minutes Low Battery Indicator: Approximately 7.2 V Statement of Conformity Greenlee Textron Inc. is certified in accordance with ISO 9001 (2000) for our Quality Management Systems. The instrument enclosed has been checked and/or cali- brated using equipment that is traceable to the National Cleaning and Storage Institute for Standards and Technology (NIST).