Pump Features
Inflation cycle
indicator
0 0 : 0 3 : 06
First 3 minutes (ramp
up) indicator
Patient compliance
counter
Patient compliance
counter reset push
button
4
Bed hanger release
Graphical display
Pump indicator lights
Telescoping bed hanger
Single / Dual Leg
operation indicator
Battery indicator
* Note: not all VenaFlow Elite
Systems will contain a battery
Single/Dual leg
operation push button
ON/OFF/RESET push
button
Cleaning Instructions
The tube assembly is reusable. Both the pump case and the tube assembly can be
cleaned with mild soap solution, antiseptic or disinfectant wipes. Avoid excessive fluids
over the top of the system. Do not submerge.
Warranty
DJO, LLC will repair or replace all or part of the 30B or 30B-B system for material or
workmanship defects for the lifetime of the system, excluding battery packs.
DJO will repair or replace all or part of the cuffs, tube units and battery packs for
material or workmanship defects, not including normal depreciation of battery life or
charge capacity, for a period of six months from the date of sale.
Optional Tube Alarm Verification Test
The Tube Alarm Verification Test is provided for facility verification purposes only and is
not required prior to use.
Attach tube assembly to pump and connect power cord to electrical outlet. Simulate
a kinked tube situation by folding the tube over onto itself once and holding it for 4-6
minutes. The second cycle after the simulated kink should activate an audible alarm
and this symbol
will appear with "CHECK TUBES" on the graphical display and the
pump indicator lights on the side of the system will flash red.