  • Page 1 Operator’s Manual Prophylaxis for Deep Vein Thrombosis for use in a clinical setting or in the home...
  • Page 2: Table Of Contents

    Table of Contents Design Philosophy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Function .
  • Page 3: Design Philosophy

    .¹ The Aircast VenaFlow Elite System leverages in the home . The VenaFlow Elite System is designed to be used only with VenaFlow the same clinically proven technology as the existing VenaFlow offering, but is Elite cuffs .
  • Page 4: Pump Features

    Pump indicator lights Call for service Telescoping bed hanger Inflation cycle indication (Bars indicate cuffs are inflating) See VenaFlow Elite System operator’s manual Safety classification: Class 1 BF 250V 2A IEC FA 5X20MM Battery Separate collection for electrical and electronic equipment. Remove and dispose of batteries in accordance with local or national requirements.
  • Page 5: Pump Set-Up

    Cuff Application 1) Hang pump from bed frame (foot of The VenaFlow Elite System will automatically detect the cuff that is attached to the bed), bed rail, or rest on floor or table . device and apply appropriate pressures . You must attach the cuffs prior to powering To use the telescoping bed hanger, on the device .
  • Page 6: Pump Operation

    S Mode Operation feature, complete the following steps: When powered on, the VenaFlow Elite System will default to standard rapid inflation 1) Start with the system powered off . mode . If a slower inflation is desired, select the S mode operation button .
  • Page 7: Compliance Alarm

    . be stored between -20~45 °C . 2) The VenaFlow Elite System with a fully charged battery will last for approximately NOTE: The battery self-discharges during storage . Higher temperatures (above 20 °C 2 hours . It takes an estimated 2 hours to fully charge the battery when the unit is or 68 °F) reduce the battery storage life .
  • Page 8: Low Battery Alarm

    Low Battery Alarm Handling Precautions 1) When there is approximately 15 minutes of charge left in the • Do not disassemble, crush, or puncture a battery. battery, an audible alarm will beep 3 times every minute . When there is less than approximately 5 minutes of charge •...
  • Page 9: Replacement Kit Instructions

    Replacement Kit Instructions Cleaning Instructions Replacement kits may be ordered through customer care . Both the pump case and the tube assembly can be cleaned with a mild soap solution, antiseptic or disinfectant wipes . DJO tested numerous disinfectants on the device and Instructions for Installing/Removing/ found no adverse effects using isopropyl alcohol and hydrogen peroxide (Oxivir Tb) .
  • Page 10: Ordering Information

    30BI-S VenaFlow Elite System, International with S Mode Bed hanger: accommodates up to 3 .5” 30BI-SB VenaFlow Elite System, International with S Mode & Battery Battery: Li-ion polymer 11 .1V, 4250 mAh Fuse: 250V 2A IEC FA 5X20MM Storage Accessories: Storage/Transportation Temperature: 0º...
  • Page 11 DJO, LLC A DJO Global Company 1430 Decision Street Vista, CA 92081-8553 • USA T 800.336.6569 • F 800.936.6569 DJOglobal .com Rx Only 30B119 Rev A...