Fluke 7038 User Manual page 38

Calibration bath
Hide thumbs Also See for 7038:
Table of Contents

Advertisement

8 General Operation
Temperature Controller
The controller allows the operator to set the bath temperature with high resolu-
tion, set the cut-out, adjust the proportional band, monitor the heater output
power, and program the controller configuration and calibration parameters.
The controller may be operated in temperature units of degrees Celsius or Fahr-
enheit. The controller is operated and programmed from the front control panel
using the four key switches and digital LED display. The controller may also be
optionally equipped with a serial RS-232 or IEEE-488 GPIB digital interface
for remote operation. Operation of the controller using the front control panel is
discussed following in Section 9. Operation using the digital interfaces is dis-
cussed in Section 10.
When the controller is set to a new set-point the bath heats or cools to the new
temperature. Once the new temperature is reached the bath usually takes 10–15
minutes for the temperature to settle and stabilize. There may be a small over-
shoot or undershoot of about 0.5°C.