Dichiarazione di conformit seguito - contrassegnati con H-606 Deluxe Bench Scale Instrument Type/Waagentyp Marked with: Gekennzeichnet mit: Munis de la mention: INSTRUCTION MANUAL Con el distintivo: Contrassegnati con la Marcatura: Model CD-11 Indicator H-606 PAGE 1 OF 30 0908 IH-606...
CD-11 Indicator MITS FOR A CLASS A DIGITAL TABLE OF CONTENTS T HARMFUL INTERFERENCE OVERVIEW OF CONTROLS AND INDICATOR FUNCTIONS ................6 S EQUIPMENT GENERATES, 1. GETTING TO KNOW YOUR INDICATOR ......................7 AND USED IN ACCORDANCE DIO COMMUNICATIONS. OP- 1.1 Introduction ...............................
Page 3
CD-11 Indicator CD-11 Indicator OVERVIEW OF CO TABLE OF CONTENTS (Cont.) 4.6 Clear Tare Operation ............................22 4.7 Unit Switch Operation ............................22 4.8 Parts Counting Operation ..........................23 4.9 Establishing a New Average Piece Weight ......................23 4.10 Returning to a Weighing Mode ........................24 4.11 Returning to a Preset APW ..........................
These factors w on a stable level surface a Thank you for deciding to purchase a CD-11 Indicator from Ohaus. The Ohaus CD-11 Indicator is a rugged, reliable, achieved when the area is electronic weight indicator designed for easy operation. The CD-11 Indicator has been NTEP tested and complies with Class III and IIIL requirements of NIST HB44 for 1:5000 performance in pounds, kilograms, or grams.
CD-11 Indicator CD-11 Indicator 2.2 Selecting the Location The Indicator should be used in an environment which is free from corrosives, vibration, temperature or humidity extremes. These factors will affect displayed weight readings. Scale bases used with the Indicator should be located on a stable level surface and kept away from vibrating sources such as large machinery.
(-) ends rest against the springs. NOTE: It is recommended that when the CD-11 is operated from batteries, the Auto-Off Timer feature be turned on to extend battery life. When setting up the Indicator, refer to Intial Setup,Readout menu, paragraph 2.7.5.
Indicator and plug the Zero/Menu Button Press and hold to enter menu. Enters menu and steps through sub-menus vertically. 2.7.2 Menu Structure The following table illustrates the menu structure in the CD-11 Indicator. MAIN MENU making sure the batteries are rEAd...
2.7 Initial Setup (Cont.) 2.7.4 Setup Menu 2.7.3 Load Cell Capacity Information The CD-11 Indicator Setu Review the specifications of the scale base to be used with the Indicator. Make sure the settings you select in the eters to match the Indicat indicator are compatible with the scale base.
2.7 Initial Setup (Cont.) 2.7.4 Setup Menu The CD-11 Indicator Setup Menu must be entered the first time the Indicator is used to set the scale base param- settings you select in the eters to match the Indicator. Do not attempt to calibrate the Indicator before setting up the Setup Menu. All other kg) table.
CD-11 Indicator CD-11 Indicator 2.7 Initial Setup (Cont.) 2.7 Initial Setup ( 2.7.5 Readout Menu 2.7.5 Readout Me The Readout menu is used to adapt the Indicator to environmental conditions, set measuring units on/off, parts counting, auto zero tracking, timer on/off and retain zero data. It contains 10 submenus. Review all of the settings available before proceeding.
Page 12
CD-11 Indicator CD-11 Indicator 2.7 Initial Setup (Cont.) 2.7.5 Readout Menu (Cont.) ring units on/off, parts Procedure Review all of the settings STABILITY (Cont.) • Press Zero/Menu button, StAb1 is displayed.The stability range specifies the weighing results and must he Readout menu, proceed...
CD-11 Indicator CD-11 Indicator 2.7 Initial Setup (Cont.) 2.7 Initial Scale S 2.7.5 Readout Menu (Cont.) 2.7.6 Print Menu Procedure The Print menu provides rate, Parity Bit, Data Leng AUTO POWER OFF • Press Zero/Menu button, AOtOFF is displayed. This is the Auto Off Timer.
CD-11 Indicator CD-11 Indicator 2.7 Initial Scale Setup (Cont.) 2.7.6 Print Menu The Print menu provides data communication settings which can be entered. It contains 6 submenus: Reset, Baud rate, Parity Bit, Data Length, Stop Bits and End/Store. Procedure , AOtOFF is displayed. This...
Lockout Switch menu (LOCSW) is a software controlled option which can lock the settings in the Calibration, Setup, CD-11 Indicators are equi Readout, and Print menus to prevent tampering. When used in conjunction with the Lock Switch (jumper) on the...
CD-11 Indicator 2.7 Initial Scale Setup (Cont.) 2.7.8 Connecting the RS232 Interface gs in the Calibration, Setup, CD-11 Indicators are equipped with a standard IBM k Switch (jumper) on the compatible, bi-directional RS232 interface for communi- and not changed by an cation with printers and computers.
3. CALIBRATION AND SEALING Legal for Trade (LFT) ope Model CD-11 Indicator requires span calibration before using. Span calibration ensures that the Indicator reads the Calibration, Setup, Re correctly within specifications using weight values of 20% to 100% of capacity. For best results, calibrate at or near menu settings to ON lock full capacity.
, -C- is displayed while the ng and then displays the • Replace the batteries and battery cover. NOTICE: The CD-11 has been tested and found to cessful, the calibration mass comply with Class lll requirements of NIST Handbook 44.
CD-11 Indicator CD-11 Indicator 4. OPERATION 4. OPERATION (C Before using the Indicator, make sure it has been properly set up and calibrated. Refer to Sections 2 and 3 and review settings. Four buttons are used: ON/OFF - Power on or OFF, PRINT/UNITS - short press prints data, long press changes unit of measure, ZERO/MENU - short press = 0, long press = enter setup menu, G/N/T/TARE - short press repeated selects UNITS for display, long press = enter tare.
CD-11 Indicator CD-11 Indicator 4. OPERATION (Cont.) o Sections 2 and 3 and hort press prints data, long 4.5 Gross/Net/Tare Recall Operation p menu, G/N/T/TARE - short When a container has been placed on the platform and tared, it's weight is stored in memory. Adding material to dicator the container is shown as NET weight.
CD-11 Indicator CD-11 Indicator 4. OPERATION (Cont.) 4. OPERATION (C 4.7 Unit Switch Operation (Cont.) If enabled in program mode, additional units of measure may be utilized beyond the primary unit of measure used for callibration or display. Switching between primary and secondary units is accomplished by pressing the Print/Units button.
CD-11 Indicator CD-11 Indicator 4. OPERATION (Cont.) 4.9 Establishing a New Average ary unit of measure used for Piece Weight (APW) (Cont.) by pressing the Print/Units ables detail the graduation • Press Zero/Menu button to accept current sample. Pcs Gross Brutto The new APW is established.
CD-11 Indicator CD-11 Indicator 4. OPERATION (Cont.) 5 CARE AND MAIN 4.12 RS232 Commands To keep the Indicator operat sary, a cloth dampened with All communication is accomplished using standard ASCII format. Characters shown in the following table are ac- knowledged by the Indicator.
CD-11 Indicator CD-11 Indicator 5 CARE AND MAINTENANCE To keep the Indicator operating properly, the housing should be kept clean and free from foreign material. If neces- sary, a cloth dampened with a mild detergent may be used. following table are ac- has not recognized the 5.1 Troubleshooting...
5.3 Service Information Load cell input sensitivit If the Troubleshooting section does not resolve or describe your problem, you will need to contact an authorized Ohaus Load cell drive Service Agent. For Service assistance in the United States, please call Aftermarket, Ohaus Corporation toll-free at Display (in/cm) (800) 526-0659.
Page 26
Load cell excitation voltage 5V dc Load cell input sensitivity Up to 3mV/V contact an authorized Ohaus Load cell drive 60 mA at 5V dc (drives up to 4 x 350 ohm load cells) s Corporation toll-free at Display (in/cm) LCD (1.0/25.4)
Page 27
CD-11 Indicator ����������������� ��������������� ������������� ����������������������������� ������������� ��������������������� ship from the date of riod Ohaus will repair, charge, provided that or misuse, exposed to ������������������ side of the product, or erly returned warranty he authorized dealer. Corporation shall not ase contact Ohaus or �...
Page 29
�������������������������� ��������������������������������� BEFORE CALIBRATION OR USE, REMOVE THE RED PLASTIC SHIPPING TABS IN THE CORNERS OF THE SCALE BASE. ������������������������� Four Overload Bolts - Factory Setup Remove Shipping Spacers Do Not Remove Screws �������������������������������������������������������������������� ��������������� ����������������������������������������������������������������������������������������� ������������������������������������������������������������������������������������� ����������������������������������������������������������������������������������� ���������� ��������������������� �����������������������������������������������������������������������������������...