Series 200
Insertion Style Flow Sensors
IMPORTANT:
READ AND KEEP FOR REFERENCE.
Installation & Operation Manual

Advertisement

Table of Contents
loading

Summary of Contents for Badger Meter 226SS

  • Page 1 Series 200 Insertion Style Flow Sensors IMPORTANT: This manual contains important information. READ AND KEEP FOR REFERENCE. 872020-EN (August 2012) Installation & Operation Manual Rev. 8...
  • Page 2: Table Of Contents

    HOT TAP INSTALLATION FOR 225BR, 226BR, and 226SS....... . .4...
  • Page 3: Introduction

    Any circumferential location is correct for installation in vertical pipes. Used in conjunction with any Badger Meter impeller flow monitor or transmitter, Badger Meter non-magnetic flow An insertion depth of 1-1/2 inches for pipe sizes sensors provide an accurate rate of liquid flow as well as 2.5 inches and larger is required for accurate flow...
  • Page 4: Alignment Of Flow Sensor

    225BR, 226BR, AND 226SS Clean O-rings and flow sensor sleeve, and lightly Badger Meter Series 200 hot tap style liquid flow sensors lubricate O-rings with silicone grease from the packet are designed for use in cases where pipelines will be in provided or some other acceptable lubricant.
  • Page 5: If The Pipe Is Depressurized And Drained

    225H consult technical bulletin DID-001. development of the flow profile. For pipe sizes 2½” and above; all Badger Meter sensors Install either a service saddle or welded pipe fitting (2 are inserted 1 1/2” from the inside wall of the pipe. The...
  • Page 6: Electrical Installation "Standard" Sensors

    Route the cable from the sensor to a Badger Meter hex mounting adapter. At this point the sleeve will flow monitor/transmitter. The cable may be extended have been inserted past the top two “O”-rings in the...
  • Page 7: Electrical Installation "High Temperature" Sensors

    ELECTRICAL INSTALLATION absolutely necessary. The Series 200 sensor is approved, as an entity, as Route the cable from the sensor to a Badger Meter intrinsically safe when installed in conformance with flow monitor/transmitter. The cable may be extended Badger Meter installation drawing 06-480-002 as specified up to 2000 feet, using 2-conductor shielded 20 AWG on the blue label identifying an intrinsically safe sensor.
  • Page 8 Series 200 Flow Sensors Page 8 August 2012...
  • Page 9: Calibration

    CALIBRATION Column 6 This column indicates the suggested flow range of Badger Meter sensors use unique K and Offset numbers sensors in each pipe size. for calibration. These numbers are derived from calibration Badger Meter sensors runs using NIST traceable instruments. Using both a K and...
  • Page 10 Series 200 Flow Sensors Column 1 Column 2 Column 3 Column 4 Column 5 Column 6 Suggested Operating Pipe Size Pipe O.D. Pipe I.D. Offset Range (GPM) 3 inch Sch 10S 3.500" 3.260" 5.009 .090 12-400 Std. Wt., Sch 40 3.5"...
  • Page 11 Installation & Operation Manual Column 1 Column 2 Column 3 Column 4 Column 5 Column 6 Suggested Operating Pipe Size Pipe O.D. Pipe I.D. K Value Offset Range (GPM) 14 inch Sch 10S 14.00" 13.500" 122.307 0.391 200-6,000 Sch 20 14.00"...
  • Page 12: Impeller Assembly And Shaft Replacement

    Refer to figure 4. To reinstall, position the impeller in the cavity oriented as in step 4 so that the impeller If you are replacing an existing Badger Meter sensor and blades point into the flow direction and toward have already calibrated your flow monitor/transmitter, the small alignment hole located beside one of the no calibration changes are necessary.
  • Page 13: Specifications

    UNS C83600 analog transmitter unit. Rated to 1050C. May be extended to a maximum of 2000 feet with similar Sensor Sleeve and Hex Adapter for 220SS and 226SS cable and insulation appropriate for application. • Series 300 stainless steel...
  • Page 14 Series 200 Flow Sensors This page intentionally left blank. Page 14 August 2012...
  • Page 15 Installation & Operation Manual This page intentionally left blank. August 2012 Page 15...
  • Page 16 The Americas | Badger Meter | 4545 West Brown Deer Rd | PO Box 245036 | Milwaukee, WI 53224-9536 | 800-876-3837 | 414-355-0400 México | Badger Meter de las Americas, S.A. de C.V. | Pedro Luis Ogazón N°32 | Esq. Angelina N°24 | Colonia Guadalupe Inn | CP 01050 | México, DF | México | +52-55-5662-0882 Europe, Middle East and Africa | Badger Meter Europa GmbH | Nurtinger Str 76 | 72639 Neuffen | Germany | +49-7025-9208-0 Czech Republic | Badger Meter Czech Republic s.r.o.