See Fig. 3.6 for models 7651, 7652, 7660 and 7661, or Fig. 3.7
models 7654, 7655, 7656, 7664 and 7665, to access the sensor
holder.
3
Thread the
connecting cable
tag ends through
the cable gland
at the top of the
cover.
1
Loosen the
gland nut.
2
Remove the
top cover by
pulling it off.
4
Unscrew and
remove clamp
ring.
5
Slide off
stem flange.
6
Remove sensor
holder from the
flow chamber.
Fig. 3.6 Access to Sensor Holder for Flow and
In-line Systems – Models 7651, 7652, 7660 and 7661
C o n n e c t i n g
Cable
Gland
Nut
Top
Cover
Stem Flange
Locating Groove
Lower
'O' Ring
Reference
Junction
3 MECHANICAL INSTALLATION...
Unscrew and remove the
3
top cover.
Drop the plug end of the
cable assembly down
through the tube and
thread the tagged end
through one of the side
4
ports.
5
Fit the gland
provided, but do
not tighten at this
stage.
2
Carefully withdraw
sensor holder.
1
Unscrew the immersion
skirt. Ensure the lower
'O' ring is not lost.
Note. If using a cleaning option refer to the
appropriate manual (IM/7600–CLN) as the sensor
holder and reference electrode are different.
Fig. 3.7 Access to Sensor Holder for Dip Systems –
Models 7654, 7655, 7656, 7664 & 7665
Side Port
Upper
'O' Ring
Lower
'O' Ring