Sterilmatic - digital electric sterilizer (14 pages)
Summary of Contents for Market Forge Industries STM-EL
Page 1
STERILMATIC ANALOG ELECTRIC STERILIZER PARTS AND SERVICE MANUAL EFFECTIVE JULY 19, 2017 Superseding All Previous Parts Lists. The Company reserves the right to make substitution in the event that items specified are not available. ERRORS: Descriptive and/or typographic errors are subject to correction. MARKET FORGE Telephone: (802) 658-6600 Fax: (802) 860-3732 www.marketforge.com...
Discard the old gasket, replace it with a new one (no cement is required), and reinstall the door assembly. Make an operational check for leakage and ad- just the door, if necessary. JULY 19, 2017 STM-E, STM-EL, STM-EX & STM-ELX STERILIZERS...
Page 4
Pivot Spring Bearing 91-2718 Right and Left Door Spring (sold as a pair) 10-1776 10-32 Machine Screw 1/2” Long 10-2666 Door Gasket 95-3204 Door & Door Spring Assembly 95-0124 Items 1 through 5 JULY 19, 2017 STM-E, STM-EL, STM-EX & STM-ELX STERILIZERS...
13.5 PSI; Center bank of elements OFF at 15.5- 15.0 PSI; STM-EL minor compensating adjustments can be made by turning The range of operation of Model STM-EL is adjusted by the adjusting screws counter-clockwise to increase pres- simply turning the adjusting knob on the outside of the sure and clockwise to decrease pressure.
Page 7
When the Timer signifies the end of the cycle, the electric current to the contactors will auto- matically be broken. Figure 6 JULY 19, 2017 STM-E, STM-EL, STM-EX & STM-ELX STERILIZERS...
Sterilmatic, replace it. The timer is replaceable only as a complete unit as factory repairs to it would not be practical in the economical inter- Figure 7 ests of the customer. JULY 19, 2017 STM-E, STM-EL, STM-EX & STM-ELX STERILIZERS...
Page 9
The expansion of the thermostatic element will make a seal against the seat to enclose the live steam within the sterilizing compartment and a steam pressure build-up will occur. JULY 19, 2017 STM-E, STM-EL, STM-EX & STM-ELX STERILIZERS...
1. Pressure cutting off at the tem- 1. See adjusting the range of opera- lbs. of pressure is reached. perature control too soon. tion and calibration of the tem- perature control. Readjust. JULY 19, 2017 STM-E, STM-EL, STM-EX & STM-ELX STERILIZERS...