5. Troubleshooting
The
Flowtron Excel
the system will sense the fault and briefly flash a message on the display.
If the same fault continues for 10 successive inflations, the audible alarm will sound
and a flashing message will remain on the display until corrective action is
completed.
The exception to this is an
Display
No display,
no operation
Alarm Cancel
Corrective Action
Check (when using
On/Off Reset)
system features an audible and visual alarm. If a problem occurs,
fault which will alarm immediately.
Problem
1. Hose disconnected at garment.
2. Garment leak.
3. Low pressure.
1. Hose kinked causing a blocked tube.
2. Hose disconnected at pump.
3. Single garment attached without
pressing 'single leg' button.
Pump failure.
1. Power failure.
2. Fuse blown.
In the unlikely event of a display electronics failure, the
red LED on the top panel will rapidly flash
(approximately four times a second). Do not use the
pump: refer to service.
After a fault has been corrected, the alarm can be
cancelled by two methods:
1. Switch the pump off, then on again, using the on/off
switch.
2. Allow the pump to run until it senses a normal
inflation; it will then reset itself.
To confirm that the fault has corrected, watch the
display for about one minute. If everything is normal,
there will be no
time.
Corrective Action
1. Check the hose connection at
garment end.
2. Check garment and replace if
faulty.
3. Call service engineer.
1. Check hoses for kinks or
obstructions.
2. Check the hose connection at
pump outlet.
3. Press 'single leg' button, if only
one garment to be used.
DO NOT USE PUMP.
Call service engineer.
1. Check power supply.
Check power cable.
2. Call service engineer.
or
message flashing during that
7
Need help?
Do you have a question about the FLOWTRON EXCEL and is the answer not in the manual?