• The pump is connected to an electrical outlet
• All tubing connections are secure
Operation Turn the power switch on which will illuminate green.
The pump performs a two-second self test cycle where
the display, LED's and alarm are tested. The compressor
then proceeds directly to the inflation cycle.
The garments will inflate alternately. The first garment
inflates for approximately 12 seconds and is deflated for
approximately 48 seconds.
The second garment inflates 30 seconds after the
inflation of the first garment and follows the same
inflation/deflation cycle.
Verify that the display is indicating the desired output
pressure. Refer to section 4 -
specific pressure setting instructions.
Pressure Adjustment
5