The angle adjustable upright trainer Instructions for use...
Page 2
Please note that guidelines and representations in these Instructions for use may deviate from your product due to differing equipment. We reserve the right to make technical modifi cations. Important information! Ensure that these Instructions for use remain with the product. Your Schuchmann Team...
Page 3
Content 1. Preparation of the upright trainer for use ..................4 1.1 Delivery ............................4 1.2 Safety measures prior to use ......................4 1.3 Safe disposal ..........................4 1.4 Where to store the Instructions for use ..................4 2. Product description ..........................5 2.1 Material information ........................5 2.2 Handling and transport .........................5 2.3 Application areas, use according to the intended purpose ............6 2.4 Use not in accordance with the intended purpose / warning guidelines ........7...
1. Preparation of the upright trainer for use 1.1 Delivery On receiving the product, please check it for completeness, lack of faults and check it for any transport damage. Inspect the goods in the presence of your forwarder Should transport damage have occurred, please arrange for an inventory (deter- mination of the faults) to be made in the presence of the forwarder.
2. Product description 2.1 Material information The base frame and the individual elements are made of steel which is non-corro- ding and powder-coated. All body supports (except for the knee pelotte pads) are padded and covered. The covers are made of 100% polyester or polyamide textile substrate, and are flame resistant (according to DIN EN 1021-1+2).
2.3 Application areas, use according to the intended purpose Indications The TIM upright trainer is designed for interior use and serves to train standing in users who can hardly stand or who cannot stand, in particular due to: • Paraplegia •...
• Therapy table with edge • 4 braked castors (75 mm) with total locking device 2.5.1 Equipment acc. HMV (Medical Aids Register) Acc. the HMV (Medical Aids Register), the upright trainer TIM requires the provision of heel edges and footstraps.
2.6 List of accessories • Chest pelotte pad with lateral guide • Knee pelotte pads (optional with (optional angular adjustment) positioner brackets) • Buttocks pelotte pad • Headrest • Ventral pelvic pelotte pad • Chin support • Pelvic guidance pelotte pad with •...
2.8 General settings User dimensions TIM is completely assembled and delivered in the smallest possible setting. Before the user is placed into the upright trainer, the neces- sary presettings must be carried out. Presettings Taking the user's dimen- Carrying out adjustments to the upright trainer...
2.9 Getting in and out Once all presettings have been carried out, ple- ase first lock all four castors for getting in and out (A). Now bring the central column into vertical positi- on (see Point 3.1) and open the swivel arm of the pelvic frame (B).
3. Settings Settings and adjustments to the product or accessories may only be made by peo- ple who have been given the necessary instructions by a medical product advisor. Please ensure that none of the user's extremities are in the respective area when making adjustments to avoid the risk of injury.
3.3 Footplate The footplate is adjustable in height and can be altered via the inner profile using the height ad- justment on the central column (see Point 3.2). To adjust the depth, loosen the star screw (A+B), which is located under the footplate. Now bring the footplate into the required depth position.
3.5 Heel edges The heel edges are mounted onto the footplate or the divided footplate. They can be adjusted in width or angle. The height or depth adjustment is undertaken via the footplate or divided footplate, see Point 3.3 or 3.4 In order to adjust the width or the angle, loosen the wing screws (A) below the (divided) footplate and bring the heel edges into the required posi-...
3.8 Pelvic frame The pelvic frame can be adjusted in height and in depth. In order to adjust the pelvic frame in height, loo- sen the wing screws (A) and bring the pelvic frame into the required position. The patient can be straightened up using the hand crank (B).
3.9 Combined spine and pelvic pelotte pad The combined spine and pelvic pelotte pad is width, height and depth adjustable. The height and the depth of the combined spine and pelotte pad can on the one hand be adjusted as a whole via the pelvic frame (see Point 3.8).
3.10 Buttocks pelotte pad The height and the depth of the buttocks pelotte pad can be adjusted as a whole via the pelvic frame (see Point 3.8). In addition, the buttocks pelotte pad can be as- sembled in three height positions on the pelvic frame.
3.12 Chest pelotte pad with lateral guide The chest pelotte pad with lateral guide is adjus- table in width and height, also with optional angu- lar adjustment. In addition, the height can also be adjusted via the height adjustment on the central column (see Point 3.2.) The chest pelotte pad support can be adjusted as a whole in height via the clamping lever.
3.14 Back pelotte pad The height and the depth of the back pelotte pad can be adjusted as a whole via the pelvic frame (see Point 3.8). The back pelotte pad (A) itself can also be adjus- ted in height, is mounted on the swivel arm of the pelvic frame and can be swivelled over this frame to the side.
3.16 Therapy tables Four therapy tables are available in total: One version without a depression and arm supports, one version with a depression, one version with arm supports and one version with a depression and arm supports. The therapy table can be adjusted in depth, height and angle.
3.16.2 Adjustable arm supports Two therapy table versions are available with ad- justable arm supports. The adjustable arm supports for the therapy table are adjustable in width. In order to adjust the width, loosen the six screws (A) on the underside of the therapy table and bring the adjustable arm supports into the requi- red position.
The covers can be removed and washed at 30° in the washing machine. Please also pay attention to our general cleaning and hygiene advice. You can find them at www.schuchmann-reha.de/TIM 4.2 Repairs Please carry out a daily visual inspection and check the upright trainer regularly for cracks, breaks, missing parts and malfunctions.
5. Technical data Dimensions - basic model Size 1 Size 2 Size 3 Body size 75 -110 cm 100 - 140 cm 120 - 160 cm Height from foot plate to tray 48,5 - 80,5 cm 71 - 95 cm 80 - 112 cm Pelvis depth 12 - 22 cm...
· info@schuchmann-reha.de · Fon +49 (0) 5405 / 909 - 0 7.3 Product version The TIM upright trainer is available in three sizes and can be supplemented through a diverse range of accessories (see Point 2.6). 7.4 Issue of the document Instructions for use TIM - Change status G;...
Need help?