CHAPTER 7
No Digital Display
Power switch is "off" => Turn it "on"
Battery is dead => Replace or recharge as appropriate
Improper Reading
Wrong cable or cable port is being used => Choose correct cable or (see chapter
4).
Selector switch is wrong position => Move to correct position (see chapter 5).
Thermocouple sensor may be defective => Test a known good thermocouple
sensor or reference standard.
Entire display or a single decimal point is flashing
Indication of low battery power => Replace with a new or freshly recharged
cell.