Electrical Specifications - Omega HH147AU User Manual

Table of Contents

Advertisement

8

Electrical Specifications

Accuracy is specified for ambient temperatures between 18 to
28°C (64 to 82°F).
Range
Resolution
±(0.05% rdg + 0.7°C / 1.4°F)
-250°C to -100°C (-418°F to -148°F)
±(0.05% rdg + 0.7°C / 1°F)
Accuracy
-100 to +1372°C (-148 to 2501°F)
0.01%+0.03°C of read value/ °C (0.06°F/ °F) outside
Temperature
+18°C to 28°C (+64°F to +82°F) specified range
coefficient
(Under -100°C (-148°F): 0.04% of read value added to K, J,
T, E, and 0.08% read value added to N).
Temperature
scale
The above specifications do not include error of thermocouple.
TYPE-K:-200°C to +1372°C (-328°F to +2501°F)
TYPE-J:-210°C to +1200°C (-346°F to +2192°F)
TYPE-T:-250°C to +400°C (-418°F to +752°F)
TYPE-E:-210°C to +1000°C (-346°F to +1832°F)
TYPE-R / S:0°C to +1767°C (+32°F to +3212°F)
TYPE-N:-150°C to +1300°C (-238°F to +2372°F)
0.1
K / J / T / E / N
1
R / S
K/J/E/T/N Type:
R/S Type:
±(0.05% rdg + 2°C / 4°F)
0°C to 1767°C (32°F to 3212°F)
ITS-90
17