Brief Introduction ......................- 1 - Features ..........................- 1 - Features of the LabSen 753 Spear pH Electrode ..............- 1 - Proper use and maintenance of the LabSen 753 Electrode ............. - 1 - Technical Specifications....................- 2 - Instrument Technical Specs....................
Brief Introduction Thank you for purchasing the Apera Instruments SX811-SS portable food pH meter. This device is an outstanding combination of advanced electrical, sensor technology and software design. The meter is equipped with LabSen753 spear pH electrode, suitable for testing meat, bread, fruit, and dairy products such as cheese, yogurt and other solid or semi-solid food sample’s pH value.
Instrument Description 3.1 LCD Display Diagram-1 (1) — Measurement mode icons (2) — Measurement value (3) — Timing storage icon. When this icon appears, the meter is in the automatic storage mode (4) — Date and time display value, and prompts of special display mode (5) —...
3.2 Keypad Diagram-2 3.2.1. Keypad operations Short press ----- <1.5 seconds, Long press ----- >1.5 seconds. 3.2.2. Turn on the meter Press to turn on the meter: LCD full display → display the measurement mode (backlight for one minute). 3.2.3. Turn off the meter Only in the measurement mode, press to turn off the meter.
In pH measurement mode, press and hold this key to change the Long press resolution repeatedly: 0.01→ 0.1pH In the mode of manual temperature compensation (MTC), when press and hold this key, the temperature value flashes, then press this key to change the temperature value, and press to confirm ...
not work. Set time 0 from parameter P4.1 to exit from the automatic storage mode. 3.4.3 Recall stored data In the measurement mode, press key to recall the last stored measuring value. See Diagram – 7: display RM icon and storage serial number. Continue pressing key and key to recall successively the stored measuring...
Page 9
Table-2 pH standard buffer series Calibration icon NIST 1.68 or 4.00 pH 1.68 or 4.00 pH 1.68 or 4.01 pH 3-Point 7.00 pH 6.86 pH 6.86 pH calibration 10.01 pH 9.18 pH 9.18 pH 4.1.5 3-Point Calibration The meter can adopt 1 to 3 points’ calibration. The 1 point must be using 7.00 pH (or 6.86 pH) buffer solution, and then choose other buffers to do 2 point and 3...
Page 10
4.1.7 Calibration reminder Preset the interval between calibrations (starting from the time when you set it), and then the meter will remind you to calibrate at the end of that interval. For detailes, please see P1.2 (6.3). When the preset time is reached, Er6 icon will be displayed at the lower right corner of the LCD (as showed in graph-8).
4.2 pH Calibration (use 3-point, USA Standard as an example) 1) Press to enter calibration mode. CAL1 icon will flash in the upper right corner of the LCD. 7.00 pH will flash in the lower right corner of the LCD, reminding you to use pH 7.00 buffer to conduct 1 point of calibration.
6) During the calibration process, press key to exit from the calibration mode. The meter can perform one-point, two-point and three-point calibration. Calibration guide icons appear on LCD. 4.3 Customized Calibration (take 1.60pH and 6.50pH calibration solution as an example) 1) Select CUS from parameter P1.1 (please refer to Section 6.3 for customer-defined solution).
Page 13
Diagram – 12 Calibration and measurement process of pH meter Submerge the After the 1 point point Press key to electrode in pH7.00 calibration is done, Press to turn calibration enter into the solution, press press key to on the meter. calibration mode.
pH electrode zero electric potential 1.Check whether pH buffer solution is correct. out of range(<-60mV or >60mV) 2. Check if electrode is damaged. pH electrode’s slope is out of range 3.The pH electrode is aged. Replace a new one. (<85% or >110%) Enter in pre-set due calibration to Press key to perform calibration or cancel due...
(b) For organic or lipid pollutant, clean the platinum surface with detergent, then wash it in distilled water, then submerge it in the soaking solution for 6 hours. (c) For heavily polluted platinum surface on which there is oxidation film, polish the platinum surface with toothpaste, then wash it in distilled water, then submerge it in the soaking solution for 6 hours.
6.3 Submenu of pH parameter setup (press key to switch) P1.1. – Select pH standard solution (USA-NIST-CUS-CH) 1. In P1.0 mode, press to enter in P1.1 mode, refer to the left diagram. 2. When press key,USA flashes,then press key, NIST flashes,etc. When parameter flashes, press to make confirmation(USA –...
6.4 Submenu of basic parameter setup (press key or key to switch) P4.1. – Adjust timing storage time 1. In mode P4.0, press key to enter in mode P4.1, refer to the left diagram: “00:”: hours (0-99), “:00”: minutes (0-59). 2.
P4.7. – Adjust date 1. Press key, “Month” flashes, then press key, “Date” flashes, then press key, “Year” flashes. When the number flashes, press key or key to adjust date, then press key to confirm. Date display: Month - Date 2.
1 — Meter serial number 2 — Stored value display area 3 — Keys Clear — press this key to clear the data Download — press this key to download the data from the meter to the computer, pH, mV, conductivity and dissolved oxygen are classified in the file.
APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS, LLC for a period of two years from the delivery (a six-month limited warranty applies to probes). This warranty does not apply to defects resulting from actions such as misuse (violation of the instructions in this manual or operations in the manner not specified in this manual), improper maintenance, or unauthorized repairs.
Need help?
Do you have a question about the SX811-SS and is the answer not in the manual?
Questions and answers