BC Biomedical ULT-2000 SERIES User Manual

BC Biomedical ULT-2000 SERIES User Manual

Ultrasound transducer leakage tester
Hide thumbs Also See for ULT-2000 SERIES:

Advertisement

ULTRASOUND TRANSDUCER
LEAKAGE TESTER
ULT-2000 SERIES
USER MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ULT-2000 SERIES and is the answer not in the manual?

Questions and answers

Summary of Contents for BC Biomedical ULT-2000 SERIES

  • Page 1 ULTRASOUND TRANSDUCER LEAKAGE TESTER ULT-2000 SERIES USER MANUAL...
  • Page 3: Calibration Interval

    LIMITED WARRANTY ..................... 4 SPECIFICATIONS ......................4 NOTES ..........................4 CALIBRATION INTERVAL To ensure the accuracy of the ULT-2000 Series, BC Group International, Inc. recommends that it be calibrated at least once every 12 months. Calibration must be done by qualified personnel.
  • Page 4 WARNING - USERS The ULT-2000 is for use by skilled technical personnel only. WARNING - USE The ULT-2000 is intended for testing only and should never be used in diagnostics, treatment or any other capacity where it would come in contact with a patient.
  • Page 5 WARNING - VOLTAGE High Voltages are generated by the ULT-2000 when running tests. Do not touch any surface that is in contact with or connected to the ULT- 2000, including the ultrasound transducer, adapter, conductivity probe, basin or liquid medium contained in the basin, as it might be at an electrical potential of 90 to 275 VAC.
  • Page 7: Notice - Symbols

    NOTICE – SYMBOLS Symbol Description Caution (Consult Manual for Further Information) Electrical Caution (Consult Manual for Further Information) Center Negative (Refers to Battery Eliminator Connector) Per European Council Directive 2002/95/EC, do not dispose of this product as unsorted municipal waste. NOTICE –...
  • Page 8: Notice - Disclaimer

    FOR INACCURACIES. NOTICE – CONTACT INFORMATION BC BIOMEDICAL BC GROUP INTERNATIONAL, INC. 3081 ELM POINT INDUSTRIAL DRIVE ST. CHARLES, MO 63301 1-800-242-8428 1-314-638-3800 www.bcgroupintl.com sales@bcgroupintl.com ULT-2000 Series User Manual Copyright © 2017 www.bcgroupintl.com Made in the USA 7/17 Rev 16...
  • Page 9 ULT-2000 SERIES ULTRASOUND TRANSDUCER LEAKAGE TESTER The Model ULT-2000 Series is a family of Microprocessor based, Ultrasound Transducer Leakage Testers. The ULT-2010 measures both the conductivity of the cleaning medium and the leakage current of the ultrasound transducer. The ULT-2020 offers the same features of the ULT-2010, plus a Meter mode and Data Logging.
  • Page 10: Standard Accessories

    ULT-2020 (METER, DATALOG) HAS ALL THE BASIC MODEL FEATURES PLUS: • METER MODE FOR EXTENDED MEASUREMENT PERIODS • PROGRAMMABLE METER SOURCE (CHALLENGE) VOLTAGE AND FREQUENCY • PROGRAMMABLE METER TIMER • DATALOG WITH STORAGE OF 99 TEST RECORDS STANDARD ACCESSORIES: BC20-21111 UNIVERSAL BATTERY ELIMINATOR BC20-41357 PC SOFTWARE (LIMIT CONFIGURATION TOOL)
  • Page 11 CONDUCTIVITY PROBES: (not included, order separately) ULT-PC-10 DUAL CONDUCTIVITY PROBE (Short) ULT-PC-15 DUAL CONDUCTIVITY PROBE (Medium) ULT-PC-20 DUAL CONDUCTIVITY PROBE (Short & Long) ULT-PC-25 FOR USE WITH GUS CLEANING SYSTEM ULT-PC-30 DUAL CONDUCTIVITY PROBE (Flexible) ULTRASOUND TRANSDUCER ADAPTERS: (not included, order separately) TRANSDUCER ADAPTER MODEL...
  • Page 12 The following is the minimal equipment needed to test the electrical safety of ultrasound transducers: 1) ULT-2000 Series Ultrasound Transducer Electrical Leakage Tester 2) Dual Conductivity Probe (See list on page 3) 3) Ultrasound Transducer Adapter (See list on page 3)
  • Page 13 for the actual ultrasound machines. The ULT-2000 is the only battery-operated handheld tester on the market today that tests according to these established protocols, which have been adopted by diagnostic ultrasound manufacturers. It tests both the upper and lower limit thresholds for electrical leakage currents. Typical electrical safety (leakage) testing of the diagnostic ultrasound transducer should occur as part of the routine cleaning and disinfecting activity that is performed between patient ultrasound procedures.
  • Page 14 During testing, the transducer is subjected to a user selectable source voltage. This voltage is typically set to the normal operating voltage of the transducer’s ultrasound machine. In North America, the source voltage is typically set to 120 VAC @ 60 Hz. For countries where the normal operating voltage is 230 VAC, the source voltage can be set to this level, at either 50 or 60 Hz, as appropriate.
  • Page 15 NOTICE – TESTING CONDUCTIVE SURFACES SUCH AS METAL CARTS OR TABLETOPS CAN CAUSE ERRONEOUS READINGS DUE TO ALTERNATE LEAKAGE PATHS FROM TRANSDUCER ADAPTERS, CABLES, ETC. ENSURE TEST IS DONE ON A NON- CONDUCTIVE SURFACE FOR BEST RESULTS.
  • Page 16: Test Process

    TEST PROCESS The ULT-2000 series completes four intermediate tests as part of the Full Test to fully evaluate the integrity of an ultrasound probe. The following is the details of each test step: Source Voltage Test - The first step is to read the actual source (challenge) voltage that will be applied during the testing to ensure that it is within range.
  • Page 17 LAYOUT This section looks at the physical layout of the ULT-2000 Series and gives descriptions of the elements. The Locking DUT Connections are Interchangeable Locking DUT Connector Locking DUT Connector – To conductive bath via Dual Conductivity Probe – To conductive bath via –...
  • Page 18 KEYS Six tactile-touch keys with audio feedback are provided for system operation: – This key turns the unit on and off. The unit will initiate with the Main Screen. – At the Main Screen, this key initiates the Full Test, which includes a Source Voltage Test, a Self Test, a Bath Test, and a transducer Probe Test.
  • Page 19 SCREENS MAIN SCREEN – The main screen indicates that the ULT-2000 Series unit is initialized and ready for testing. This screen displays after power-up initialization, and can be accessed by pressing the key at any time other than during setup mode.
  • Page 20 BATH TEST SCREEN – This screen is accessed from the Main Screen by pressing the button. It displays Bath Test information, measurements, and progress. Bath Test Process: DUT Information Manufacturer: Model SOURCE VOLTAGE TEST - verifies the voltage circuit to be used during testing.
  • Page 21 METER MODE SCREEN (ULT-2020 ONLY) – The meter mode screen allows extended leakage current measurements, which can be useful when troubleshooting ultrasound transducer probes and cables. This screen displays the current leakage current reading and user selectable Source Voltage, Source Frequency, and Output Control. This screen is accessed from the Main Screen by pressing the key.
  • Page 22 DATALOG SCREEN (ULT-2020 ONLY) – The Datalog Screen displays the test results of up to 99 test records. The Datalog Screen includes a Record Number, Date / Time Stamp, Probe Information, Test Results, Test Measurements, and Test Limits. This screen is accessed from the Meter Screen by pressing the key.
  • Page 23 Use the key to print the displayed record to the serial printer. Below is a sample print.
  • Page 24 DEVICE CONFIGURATION SCREEN – The Device Configuration displays the current test configuration. Device Configuration parameters include Test Limit Mode, Probe Manufacturer, and Probe Model. This screen is accessed from the Datalog Screen by pressing the key. The parameters are selected by pressing the key until the desired parameter is highlighted.
  • Page 25 LOW BATTERY – When the battery life is 10% or less, the LOW BATTERY message box appears and indicates the remaining battery life. BATTERY ELIMINATOR INPUT – A 2.1 mm receptacle is provided for the 10 VDC Battery Eliminator (BC20-21103, BC20-21106) that may be used for continuous run applications.
  • Page 26: System Configuration

    SETUP Two user selectable setup menus are provided, SYSTEM CONFIGURATION and USER TEST CONFIGURATION. Enter a setup menu by pressing and holding the until the Access Code Screen appears (5 sec). The arrows are then used to enter the access code. DEFAULT ACCESS CODES SYSTEM CONFIGURATION = 1 USER TEST CONFIGURATION = 2...
  • Page 27 System Configuration Parameter Description Options Determines whether the test measurements (actual PASS / FAIL or test readings) are shown in the test screen, or if the Test Mode unit will simply give a PASS or FAIL result. Numerical Default setting is Numerical. Determines the settings for the test limits.
  • Page 28 USER TEST CONFIGURATION Below is the typical User Test Configuration Screen, followed by a table of the available parameters and a brief description of each option. User Test Configuration Parameter Description Range The Source Voltage applied during a test when Test Voltage 90-275 VAC Device Configuration is set to User Test Limits.
  • Page 29 PC SOFTWARE The included PC Software allows the user to create custom limit configurations including Manufacturer, Model, Leakage Limits, Test Voltage, and Frequency. Install and launch the PC Software, then connect the PC and ULT-2000 via cable BC20-41361 (See OPTIONAL ACCESSORIES at the beginning of this manual). Once connected, the software allows the user to download/read custom configurations to/from the ULT-2000.
  • Page 30: Communication Protocol

    COMMUNICATION PROTOCOL The communication protocol provides a means to completely configure and use the ULT- 2000 from a PC, allowing hands free or automated operation. Communication Port The Serial port is configured as 115,200 Baud Rate, 8 Data Bits, 1 Stop Bit, and No Parity. Command Syntax The command description is broken into columns;...
  • Page 31 ULT-2000 Communication Command Summary Subnodes Keywords Nodes Values OUTput VOLTage 90-275 FREQuency 50 hz, 60 hz TEST MODE NUMerical, PASSfail LIMits FIXed, DEFaults, CUSTom DMANufacturer 1-5 (Selects Manufacturer 1-5) DMODel 1-20 (Selects Model 1-20) CONFigure LIMits CONDuctivity 0 - 500 (uA) ULEAkage 0 - 500 (uA) LLEAkage...
  • Page 32 ULT-2000 Communication Command Summary (cont.) Keywords Nodes Values Value Definition Test Running Bath Only Test Full Test Test Failed Test Passed STATus? 1024 2048 Program Mode 4096 Meter Mode 8192 14 16384 15 32768 Calibration Mode VTOLerance 0-25 % FVOLtage 90 - 275 VAC FCONductivity 0 - 500 uA...
  • Page 33: Manual Revisions

    MANUAL REVISIONS Revision # Program # Revisions Made Rev 01 DT7331CA Origination Rev 02 DT7331CA Misc. Edits Rev 03 DT7331CA Adapter Information Updated Rev 04 DT7331CB Model Information Updated Rev 05 DT7331CC Dynamic Device Configuration Screen Rev 06 DT7331CC Edits to Euro Transformer, Conductivity Probe and Transducer Adapter List Updated, and Misc.
  • Page 34: Specifications

    SPECIFICATIONS SOURCE, LEAKAGE AND CONDUCTIVITY SOURCE 90 - 275 VAC, ± 1% FS (CHALLENGE) 500 µA Max Load VOLTAGE SOURCE (CHALLENGE) 50 or 60 Hz, ± 0.5 Hz FREQUENCY 0.50 - 10.00 µA, ± 0.5 µA LEAKAGE 10.0 - 250.0 µA, ± 1% Range CURRENT 250.0 - 500 µA, ±...
  • Page 35 ELECTRICAL & MISC. 9V Lithium Battery BATTERY (ANSI/NEDA 1604LC or equivalent) 10 VDC, 300mA BATTERY ELIMINATOR BC20-21111 (Universal) > 100 Full Tests CONTINUOUS (Note: Backlight set to OFF) BATTERY LIFE 1 year BAUD 115200 DATA BITS START BITS STOP BITS PARITY none HANDSHAKING...
  • Page 36 NOTES...
  • Page 37 NOTES...
  • Page 38 NOTES...
  • Page 40 BC GROUP INTERNATIONAL, INC. 3081 ELM POINT INDUSTRIAL DRIVE ST. CHARLES, MO 63301 1-800-242-8428 1-314-638-3800 www.bcgroupintl.com sales@bcgroupintl.com ULT-2000 Series User Manual 7/17 – Rev 16 LM-7331X-UM Copyright © 2017 Made in the USA...