■ Alarm/Safety Functions
Types
Auto set
Temp. deviated ±2.5℃ from set
upper/lower
value
limit alarm
No key operation for about
Auto return
45 sec. in each setting mode
Key lock switch
Key lock becomes ON
Temp. sensor
Temp. sensor open or short circuit
abnormal
Triac open circuit
Triac abnormal
Triac short circuit
Relay short circuit
Relay
Relay open circuit, heater short
abnormal,
etc.
circuit or temp. around PCB is
65℃ or more
During
OFF
Overheat
protection
During
ON
● Buzzer sound during overheat protection cannot be stopped by buzzer stop key (BZ). Turn off the main switch.
■ Performance Specification
Product Name
Model
Temperature control
range
Temperature distribution
Temperature fluctuation
range
Maximum current
Maximum power
consumption
Maximum heat
discharge
Applicable environment
condition
- Each data is measured based on our standards.
- Specification can be changed for improvement without notification.
Condition
Abnormal heating is
detected, safety
circuit operates by
independent temp.
sensor
MIR-H163
Room temp. +5℃~60℃ (AT: 0℃~ less than 20℃, no load)
Room temp. +5℃~80℃ (AT: 20℃~35℃, no load)
±1℃ (set: 37℃, AT: 20℃, no load)
±0.2℃ (set: less than 60℃, AT: 20℃, no load)
±0.5℃ (set: 60℃~80℃, AT: 20℃, no load)
2.0 A
200 W
720 kJ/h
Temperature: 0℃~35℃, Humidity: 80%R.H. or less
-3-
Display
Alarm lamp lighting
Temp. display
blinking
Display current value
―――
Alarm lamp lighting
E01 blinking
Alarm lamp lighting
E02 blinking
Alarm lamp lighting
E03 blinking
Alarm lamp lighting
E04 blinking
Alarm lamp lighting
E05 blinking
Display no change
Alarm lamp lighting
E05 blinking
Heated Incubator
Buzzer
Action Taken
After 3sec.
――
intermittently
Finish
――
setting mode
Not accept any
――
key input except
call key
Intermittently
Stop operation
Intermittently
Stop operation
Intermittently
Stop operation
Intermittently
Stop operation
Intermittently
Stop operation
Stop operation
Continuously
forcibly by
external circuit
Continuously
Stop operation
(Intermittently
forcibly by
when temp.
external circuit
drops)
MIR-H263
3.0 A
300 W
1080 kJ/h