Fault
Error code -1026
Error code 1100
Error code 1104
Error code 1116
Error code 1117
Error code 1118
Error code 1121
Error code 1153
Error code 1154
Error code 1160
Error code 1170
Error code 1172
Error code 10000
Error code 10009
2142100002L02 1611V004
Probable cause
Incorrect acquisition mode
Permitted time for scan process
exceeded
Erasure unit fault
Drive feed blocked
Feed position error
Input unit cover open
Input unit fixing mechanism miss-
ing
Unit fault
Internal data communication fault
Final radiation deflector rotation
speed not attained
SOL sensor timeout
Fault on the laser, SOL sensor or
radiation deflector assembly
Fault on the laser, SOL sensor or
radiation deflector assembly
Unit exposed to too much light
Internal communication error
warning; unit remains ready for
operation
Troubleshooting
Solution
Select a different acquisition
i
mode
Inform a service technician.
i
Update the firmware.
i
Reset the scanning modes
i
to the factory settings via the
unit interface or the Imaging
Software.
Inform a service technician.
i
Check the belt drive.
i
Check for blockage, remove
i
image plate from unit.
Inform a service technician.
i
Replace the erasure unit.
i
Remove the blockage.
i
Inform a service technician.
i
Inform a service technician.
i
Check the feed (ease of move-
i
ment, drive belts).
Close the cover.
i
Deactivate cleaning mode.
i
Insert the fixation mechanism.
i
Deactivate cleaning mode.
i
Switch the unit off and back on
i
again.
Update the firmware.
i
Switch the unit off and back on
i
again.
Update the firmware.
i
Inform a service technician.
i
Update the firmware.
i
Replace the radiation deflector
i
subassembly if the problem oc-
curs regularly.
Inform a service technician.
i
Update the firmware.
i
Inform a service technician.
i
Update the firmware.
i
Darken the room.
i
Turn the unit so that no light can
i
fall directly into the entry slot.
Update the firmware.
i