Pack Contents List Specification Instrument Layout Display symbols Maintenance Cleaning Operation Breath Test Settings Sounder User Profiles Breath-hold Countdown Timer Date and Time System Information Zero/Calibration Patient Information Troubleshooting Returns Procedure Spares and Warranty www. . covita.net | (800) 707-5751...
Micro Operating Manual Intended Use Smoking status The Micro Smokerlyzer is a breath carbon ® The cut-off between smoker and non- monoxide monitor intended for multi-patient smoker is likely to be 6ppm CO. use by healthcare professionals in smoking Micro uses a coloured traffic light system cessation programmes and as an indicator to motivate the patient to give up smoking.
● The disposable cardboard mouthpieces or on the order of a physician. are single-use only as re-use can increase the risk of cross-infection. ● See Bedfont’s Infection Control and Maintenance Guidelines for further information on infection control. www.covita.net | tel (800) 707-5751...
Micro Operating Manual Quick Start Guide Press the on/off button until the display becomes active. Release the button. Insert the D-piece into the instrument and fit a new cardboard mouthpiece. Touch the icon to start a breath test. This starts the breath-hold countdown.
The Micro+ complies with the directive EN60601-1-2 electromagnetic compatability but can be affected by cel- lular phones and by electromagnetic interference exceedding the levels specified in EN50082:1. This equipment should be moved if necessary to avoid interference. *Whichever is greater. www.covita.net | (800) 707-5751...
Micro Operating Manual Pack Contents List Instrument Operation Instructions Box of D-pieces Disposable Flatpak cardboard mouthpieces 3 x AA batteries Infection control and maintenance guidelines Warranty and registration card USB cable CD with COdata+ software Instrument Layout Display On/Off switch Aperture for D-piece Breath sampling D-piece Exhaust port for breath...
28. Breath hold timer setting: 10. Patient data: 29. Set date/time: 11. Patient data in table format: 30. System info screen: 12. Patient data in chart format: 13. Inhale: 31. Contact Bedfont or distributor for help: 14. Breath-hold and countdown: www.covita.net | (800) 707-5751...
Micro Operating Manual Maintenance Cleaning Wipe the instrument and external D-piece Calibrate in accordance with procedure on surfaces with a product specifically page 13. developed for this purpose. Bedfont provides an Instrument Cleansing Wipe. Replace batteries when indicated. Never use alcohol or cleaning agents Replace D-piece every month or sooner if containing alcohol or other organic solvents visibly soiled or contaminated.
28 days. ● The main menu displays three symbols: Breath test Settings Patient select Note: The patient select icon will not be accessible until patient data has been downloaded from COdata and will remain greyed out. www.covita.net | (800) 707-5751...
Micro Operating Manual Breath Test ● Attach a D-piece and a new cardboard ● To view the corresponding %FCOHb mouthpiece to the Micro . Check all value, touch the . To go back to the connections are pushed firmly together. ppm and %COHb reading, touch the again.
Touch to access system information such as software version. Touch return to previous menu. R.West; Effect of duration of breath holding on expired air carbon monoxide concentration in cigarette smokers; Addictive www.covita.net | (800) 707-5751 Behaviours, Vol 9 (1984), 307-309...
Micro Operating Manual Zero/Calibration ● The instrument should be calibrated at ● Touch to begin calibration. the temperature at which it is expected ● Immediately open the fine control valve to be used. See Warnings page 4. by turning the control knob anti-clockwise and allow the gas to ●...
COdata Touch to highlight the required patient. Touch to view data in table format or to view in graphical format. Touch to return to previous menu. www.covita.net | (800) 707-5751...
Micro Operating Manual Troubleshooting ● If, after an attempted calibration with ● If the unit fails to turn on properly, or 50ppm CO gas, the display shows if the low battery symbol is showing, , the gas value was not within the replace the 3 x AA alkaline batteries.
● If the monitor is still in warranty, Bedfont will repair it and return it with an Engineer’s Report, free of charge. If the monitor is found to simply require calibrating, a fee will be charged. www.covita.net | (800) 707-5751...
Micro Operating Manual Spares Spares: D-pieces Disposable Flatpak mouthpieces Calibration gas and kits Instrument cleansing wipes AA alkaline batteries The above spares are available from Bedfont Scientific Ltd, UK. For spares availability in all other countries contact your local distributor. It is recommended that only Bedfont spares are used.