STORAGE CONDITIONS
Temperature range
Humidity range
Pressure range
Maximum time
SENSOR SPECIFICATIONS
Gas Name
Abbr.
Oxygen
O
2
Carbon Monoxide
CO/H
with low H
Low
2
interference
Carbon Monoxide
CO
Hydrogen Sulfide
H
S
2
Nitrogen Dioxide
NO
2
Sulfur Dioxide
SO
2
Combustible
LEL
Methane
CH
4
*The accuracy specification for each sensor is ± the stated percentage or 1 unit of resolution, whichever is greater.
TOXIC GAS SENSOR CROSS-SENSITIVITY TABLE
Target Gas
CO
CO
100
H2S
5
SO2
0
NO2
-5
Cl2
-10
ClO2
—
HCN
15
HCl
3
PH3
—
NO
25
H2
22
NH3
0
47
0−25 ºC (32−77 ºF)
40–70% relative humidity (RH) noncondensing
0.9–1.1 atm
Up to 6 months
Note: Industrial Scientific recommends that infrequently used lithium-ion batteries be
fully charged every four months.
Measuring
Resolution
Range
0–30% vol
0.1% vol
0–1000 ppm
1 ppm
2
0–1000 ppm
1 ppm
0–500 ppm
0.1 ppm
0–150 ppm
0.1 ppm
0–150 ppm
0.1 ppm
0–100% LEL
1% LEL
0–5% vol
0.01% vol
CO
(H2
Low)
H2S
SO2
NO2
100
1
1
0
5
100
1
-40
5
5
100
0
5
-25
-165
100
0
-20
-25
10
—
—
—
—
—
10
50
1
—
0
5
0
—
—
—
—
40
1
1
5
3
0.3
0.5
0
0
0
0
0
Ventis™ MX4 Product Manual
Accuracy at Time and
Temperature of
Calibration*
± 0.5%
± 5% (0-300 ppm)
± 15% (301-1000 ppm)
± 5%
± 5%
± 10%
± 10%
± 5%
± 5%
Sensor
Cl2
ClO2
HCN
HCl
0
0
0
0
-3
-25
10
300
0
0
—
40
45
—
-70
—
100
60
-20
6
20
100
—
—
0
0
100
35
2
0
0
100
—
-100
425
300
—
—
-5
—
0
0
0
0
0
0
0
0
Response
Response
Time
Time
(typical)
(typical)
T50
T90
15
30
8
17
15
50
15
30
10
30
20
80
15
35
15
35
PH3
NO
H2
NH3
0
0
20
25
10
20
—
0
0
—
30
0
-20
0
0
—
—
—
1
0
30
0
15
0
100
—
—
—
100
30
0
0
100
0
0
0
100
© 2014 Industrial Scientific Corporation
0
25
-40
-10
-50
—
5
0
—
0
0