Ventis™ MX4 Product Manual
Bump Test Percentage Requirement Screen
Value range: 50% to 99%
Value increment: one percent
Bump Test Response Time Screen
Value range: 30 to 300 seconds
Value increment: five seconds
Latch Alarm Set Screen
Options
0 = Normal mode
1 = Latching mode
Zero In-field Screen
Options
0 = Disable/off
1 = Enable/on
Calibration In-field Option Screen
Options
0 = Disable/off
1 = Enable/on
Calibration Due Alarm Option Screen
Options
0 = Disable/off
1 = Enable/on
© 2014 Industrial Scientific Corporation
Sets the percentage of calibration gas the monitor expects to be
exposed to.
Press ENTER to edit the value, if needed; press repeatedly or
hold down to speed the increment pace.
Press ON/OFF/MODE to set the value and advance to the Bump
Test Response Time Screen.
Sets the bump test response time period.
Press ENTER to edit the value, if needed; press repeatedly or
hold down to speed the increment pace.
Press ON/OFF/MODE to set the value and advance to the Latch
Alarm Set Screen.
When enabled, if the monitor goes into any gas-related alarm, it
will remain in alarm until after the gas concentration is less than (or
more than for oxygen) the alarm set point, and the monitor user
presses the ENTER button for one second.
Press ENTER to edit the value, if needed.
Press ON/OFF/MODE to set the value and advance to the Zero
In-field Screen.
When enabled, all monitor users are permitted to zero the monitor
from the gas monitoring mode.
Press ENTER to edit the value, if needed.
Press ON/OFF/MODE to set the value and advance to one of two
screens.
If Zero In-field is enabled, the Calibration In-field Option
screen displays.
If Zero In-field is disabled, the Calibration Due Alarm screen
displays.
When enabled, all monitor users are permitted to calibrate the
monitor from the gas monitoring mode.
Press ENTER to edit the value, if needed.
Press ON/OFF/MODE to set the value and advance to the
Calibration Due Alarm Option.
When enabled, the monitor will activate the calibration due alarm,
in gas monitoring mode, when any sensor is due for calibration. A
flashing gas cylinder and gas type will appear on the display and
three beeps will sound every 30 seconds.
Press ENTER to edit the value, if needed.
Press ON/OFF/MODE to set the value and advance to the
Calibration Due Set-point screen.
22