Configuration; Introduction; Instructions - Ventis MX4 Product Manual

Table of Contents

Advertisement

Ventis™ MX4 Product Manual
Countdown Screen
Displays the 20 second countdown, one second at a time, from 20
to one. Options :
Enter gas monitoring mode
Enter configuration mode
Power-Off Screen
The screen displays a five-second countdown accompanied by five
beeps and LED flashes.

CONFIGURATION

Introduction

Before first time use of the monitor, its settings should be reviewed and, if needed, be adjusted. Qualified safety
personnel should complete the following tasks.
Review the monitor settings for compliance with company policy and any applicable regulations, laws, and
observed guidelines as issued by regulatory agencies and government or industry groups.
Determine which settings, if any, require adjustment.
Make the adjustments or supervise other qualified personnel in the process.
When using instrument-compatible Industrial Scientific docking station software (e.g., iNet, Docking Station
Server Administrative Console [DSSAC], or Accessory Software), maintain the mandated settings through
the software or by manually configuring the instrument settings after docking.
Monitor settings should be reviewed regularly and adjusted as needed. The following settings are adjustable or
"configurable" for the Ventis MX4.
LEL Type
Calibration Mode Setting
Low Alarm Settings
High Alarm Settings
TWA Alarm Settings
TWA Interval Settings
STEL Alarm Setting
Calibration Gas Settings
Clock Settings
Date Settings
The Ventis MX4 can be configured manually as instructed below. Any changes made take effect immediately upon
exiting the configuration mode.

Instructions

NOTICES
 The configuration mode should be accessed only by safety personnel authorized to change monitor settings
based on company policy.
 Read ALL requirements and instructions outlined below, including the screen-by-screen process description,
before beginning the configuration process.
© 2014Industrial Scientific Corporation
To enter gas monitoring mode:
allow the countdown to complete and advance
to the Gas Monitoring Screen. Proceed to the
manual section, Monitor Use and Service.
To enter configuration mode:
simultaneously press ON/OFF/MODE and
ENTER, hold for three seconds, and release.
Press ON/OFF/MODE, hold for the full five
second countdown to zero, and release to
power-off the monitor.
Display Mode Setting
Confidence Indicator (on/off)
Confidence Indicator (type)
Bump Test In-field
Bump Test Due Warning
Bump Test Time Set-point
Bump Test Percentage
Bump Test Response Time
Alarm Latch Set
Zero In-field
Calibration In-field
Calibration Due Alarm
Calibration Due Set-point
Security Code
Language Selection
Always-on Setting
Shutdown In Alarm Setting
Alarm on Dock Setting
16