ORBECO SC400 Instruction Manual

Colorimeter copper

Advertisement

SC400 Colorimeter
Copper
Instruction Manual

Advertisement

Table of Contents
loading

Summary of Contents for ORBECO SC400

  • Page 1 SC400 Colorimeter Copper Instruction Manual...
  • Page 2: Operation

    Operation Switch the unit on using the ON/OFF key. The display shows the following: Attention: The unit is designed to measure copper either with tablet reagents (Cut) or with powder packs (CuP). Please select the required method as described below. Select the test required using the MODE key: Mode Cut →...
  • Page 3: Technical Data

    Technical Data Light source: LED, Filter (l = 528 nm) Battery: 9 V-block battery (Life 600 tests), without display light. Auto-OFF: Automatic switch off 7 minutes after last keypress Ambient conditions: 5-40°C 30-90 % rel. humidity (non-condensing). DIN EN 55 022, 61 000-4-2, 61 000-4-8, 50 082-2, 50 081-1, DIN V ENV 50 140, 50 204 Method notes Prior to measurement ensure that the sample is suitable for analysis...
  • Page 4 Copper 0.05 - 5 mg/L with tablets (Cut) (a) free copper Perform zero calibration (see “Operation”). 0.0.0 Add one COPPER No. 1 tablet straight from the foil to the 10 ml water sample, and crush using a clean stirring rod. Allow to dissolve completely, close the vial, and position with ∇...
  • Page 5 Copper, free 0.05 - 5 mg/L Cu with Powder Packs (CuP) Perform zero calibration (see “Operation”). 0.0.0 Add one VARIO Cu 1 F10 powder pack straight from the foil to the water sample. Close the vial with the cap tightly and swirl the vial several times to mix the contents (Note 3).
  • Page 6: Menu Selections

    Menu selections Press the MODE key and hold. Mode Switch the unit on using the ON/OFF key. Allow the 3 decimal points to be displayed before releasing the MODE key. The “!” key allows for selection of the following menu points: ▲...
  • Page 7: User Calibration

    User calibration Note: user calibration (Display in calibration mode) factory calibration (Display in calibration mode) Press MODE key and hold. Mode Switch the unit on using the ON/OFF switch, Release the MODE key after approx. 1 second. Select the required menu points using the "!" key. After confirming the selection with the MODE key the instrument will show CAL/Cut.
  • Page 8: Resetting The Factory Calibration

    Resetting the factory calibration Resetting the user calibration to the oriinal factory setting resets all parameters. Store Date RESULT ▼ A user calibrated method will be indicated by an arrow in the display. Time To reset the calibration to the factory setting: Press both the MODE and ZERO/TEST key and hold.
  • Page 9: Error Codes

    Error codes General error codes Light absorption too great. Reasons: dirty optics. EOOI Measuring range exceeded or excessive turbidity. Result below the lowest limit of the measuring range. LO BAT Replace 9 V battery, no further tests possible. Calibration factor "out of range" EOIO Individual error codes Cut:...
  • Page 12 Orbeco-Hellige Inc., 185 Marine Street, Farmingdale, NY 11735 Phone: 631-293-4110, Fax: 631-293-8258, service@orbeco.com, www.orbeco.com...

Table of Contents