C
ALIBRATE
The instrument alarms are deactivated during the calibration to save battery life.
If "Calibrate" is selected, the instrument displays the confirmation screen shown
below. If "Cancel" is selected, the user is returned to the gas-monitoring display
screen.
If the user selects "OK", all the installed sensors are zeroed first (following the
"Zero All" steps outlined above) and then calibrated. After the zero, the results
are shown for 5 seconds and then the calibration of the first sensor begins.
The screen to alert the user to connect gas to the instrument is then shown.
Once the sensor starts to read gas, the calibration begins. The user has 5
minutes to apply gas before the calibration times out. Gas should be applied at a
flow rate of 0.5 lpm. If the user chooses to "Skip" a sensor, the instrument will
move to the next sensor. The "Abort" option aborts the calibration and shows the
"Calibration Complete" screen.
When the calibration is finished, the following screen shows the passed,
marginal, skipped, and failed sensors, when six sensors are installed.