HT-7120 Troubleshooting This Troubleshooting Guide walks you step-by-step through the diagnosis of HT-7120 customer problems. It is important to follow these steps carefully, to ensure that the problem is diagnosed correctly, and therefore the appropriate repair action is taken. Getting Started 1.
Page 4
If there is a kink, control isn’t working cut the kinked portion of the tubing off, then reconnect the tubing to the fitting. If this does not solve the problem, the air pump is not working. Replace the air pump. HT-7120 Troubleshooting Guide...
Page 5
2. Using a small screwdriver or a coin, turn the fuse holder (located on the power panel) approximately 1/8 turn counter-clockwise. The fuse holder pops out. 3. Replace it with a new fuse of the same rating. HT-7120 Troubleshooting Guide...
Press the Rolling button on the remote control. If the rolling massage works for approximately 30 My Rolling button is seconds then stops, the rolling motor sensor is not working not working properly. Replace the massage mechanism. HT-7120 Troubleshooting Guide...
Page 7
You must press and hold the width buttons to adjust the width setting. Some LEDs may not light as the width is adjusted to the new setting, however when you release the width button the LEDs indicate the current setting. HT-7120 Troubleshooting Guide...
2. Verify that the device being plugged into the convenience outlet works when plugged into a wall outlet. 3. Turn the power switch to OFF and unplug the chair from the wall outlet. HT-7120 Troubleshooting Guide...
Page 9
6. Reinsert the fuse holder into the power panel, then turn it 1/8 turn clockwise to secure it. 7. Plug the chair back into the wall outlet and turn it on. 8. If the convenience outlet is still not receiving power, replace the backrest cover. HT-7120 Troubleshooting Guide...
Page 10