Removing the Seat and the
Batteries Cover (2) ????
Replacing the Batteries
Cover(2)
Batteries
Tires
Cleaning
Turn the Seat by using the lifting Lever so that the Lever will
point 45º Right or Left. The Seat can be removed only at this
position.
Remove the Seat (7): Hold the Seat in your two hands – one
hand on the backrest and a second hand at the front lower
part of the seat and lift the seat up from its pivot.
Warning! The seat weight is 21 Kg (44lbs).
Keep lifting with your Knees/legs and not loading your Back.
Release Battery Cover (2) securing screws (2.6).
Lift the Battery Cover (2) until it is released from the Seat
Pivot .
The replacement of the Cover is done in reverse order;
MAKE SURE that the Cover (2) fits onto it's place. Secure
with the 4 screws (2.6).
Batteries weight each ~14 Kg (~30 lbs) and over.
Lift each battery only with the appropriate handle. Keep
lifting with your Knees/legs and not loading your Back.
Correct air pressure in the tires is essential for optimal
steering and stability of the Breeze C. Check air pressure
every two weeks.
Inflate to the proper air pressure: 25 ±2 psi. Breeze 4W front
tires: 25±2 psi. Golf wheels tires : 25±2 psi.
Use only a damp cloth and mild detergent. Never use a hose
for cleaning. This may severely damage the power and
electronic components.