Summary of Contents for Iso-Tech ICM 134

  • Page 1 Instruction Manual ICM 134 Clamp Meter...
  • Page 2: Safety Information

    Safety Information Read and understand this Instruction Manual completely before using this instrument. Failure to observe the warnings and cautions in this Instruction Manual may result in injury or death, or damage to the instrument and other equipment or property. If this instrument is used in a manner not specified in these instructions, the protec- tion provided by the instrument may be...
  • Page 3 - Wear suitable Personal Protective Equipment when working around or near hazardous Live conductors which could be accessible. - Do not use the thermocouple to measure the temperature of Hazardous Live conductors or equipment. Caution - Disconnect the test leads from the test points before changing the position of the function rotary switch.
  • Page 4: Maintenance

    Maintenance Do not attempt to repair this Instrument. It contains no user-serviceable parts. Repair or servicing should only be performed by qualified personnel. This instrument should be calibrated yearly, or more frequently if used in harsh conditions or if it is suspected of being inaccurate. For calibration and repair contact RS Components - the address is given at the end of these instructions.
  • Page 5: Power On/Off

    Power On / Off Auto Power Off To conserve battery power, the Auto Power -off function will turn the instrument off approximately 10 minutes after the last operation. 10m in To disable the Auto Power-off function: Turn the instrument off, press and hold the “Backlight”...
  • Page 6 ACV / DCV Resistance EN - 6...
  • Page 7: Diode / Continuity

    Diode / Continuity EN - 7...
  • Page 8 DC µA Capacitance Note – To improve the measurement accuracy of a small value capacitor, record the reading with the test leads open circuit, then substract the residual capacitance of the Instrument and leads from the final measurement. UNKNOWN MEASUREMENT RESIDUAL EN - 8...
  • Page 9 Temperature Data Hold EN - 9...
  • Page 10 CAT.Ⅲ.600V with respect to earth for the jaw. Tactile Barrier for hand guard. Do not hold the meter across the Tactile Barrier. I + (-I) = 0 Back Light EN - 10...
  • Page 11: Specifications

    Specifications 1-1 General Specifications LCD display digits : digit large scale Liquid Crystal Display. Display resolution : 4000 counts. Measurement rate : 1.5 times / sec. Overrange display : “OL” is displayed for "Ω" functions, The real value is shown for the "A" ,"V" and "°C/°F "...
  • Page 12 Measurement Category lll is for measure- ments performed in the building installation. Examples include measurements on distribution boards, junction boxes, socket- outlets and wiring and cables in the fixed installation. Measurement Category lV is for measure- ments performed at the source of the low-voltage installation.
  • Page 13: Electrical Specifications

    1-3 Electrical Specifications Accuracy is ±(% reading + number of digits) at 23°C (73.4°F) ± 5°C (± 9°F) < 80%RH. Temperature coefficient : Add 0.2 x (Specified accuracy) / °C (1.8°F), < 18°C > 28°C (< 64.4°F >82.4°F). DC / AC Volts Range DC Accuracy AC Accuracy...
  • Page 14 Diode Check and Continuity Resolution Accuracy ±(1.5%+5 dgt)* 10 mV * For 0.4V ~ 0.8V Max.Test Current : 1.5mA Max. Open Circuit Voltage : 3V Overload Protection : 600Vrms Continuity : The built-in buzzer sounds when the resistance is less than approxi- mately 100Ω.
  • Page 15: Battery Replacement

    Temperature Function Range Accuracy -40.0 °C ~ 0.0°C 1% ± 4 °C °C 0.0 °C ~ 400.0 °C 1% ± 3 °C 1% ± 8 °F -40 °F ~ 32 °F 1% ± 6 °F 32 °F ~ 750 °F °F 2% ±...
  • Page 16: Limited Warranty

    Limited Warranty This instrument is warranted to the original purchaser against defects in material and workmanship for 3 year from the date of purchase. During this warranty period, RS Components will, at its option, replace or repair the defective unit, subject to verifica- tion of the defect or malfunction.
  • Page 17 Asia Iso-Tech 460 Alexandra Road, #15-01A PSA Building Singapore 119963 7410 Pebble Drive Fort Worth Texas 76118-6961 Europe Iso-Tech PO Box 99 Corby Northamptonshire NN17 9RS United Kingdom Canada 1701 Woodward Drive Ste 108 Ottawa Ontario K2C 0R4, Canada Japan...