Hach APA 6000 Operation Manual page 9

High and low range copper process analyzer
Table of Contents

Advertisement

Figure 1
Plumbing Diagram for a Copper APA Analyzer
Item
Description
1
Tefzel tubing, 0.03 ID x 0.062 OD
2
Tefzel tubing, 0.03 ID x0.062 OD
3
Blank port
4
Blank port
5
Tefzel tubing, 0.03 ID x 0.062 OD
6
Tefzel tubing, 0.03 ID x 0.062 OD
7
Tefzel tubing, 0.03 ID x 0.062 OD
8
Tefzel tubing, 0.03 ID x 0.062 OD
9
Tefzel tubing, 0.03 ID x 0.062 OD
10
Tefzel tubing, 0.03 ID x 0.062 OD
11
Tefzel tubing, 0.03 ID x 0.062 OD
12
Tefzel tubing, 0.03 ID x 0.062 OD
13
Tefzel tubing, 0.03 ID x 0.062 OD
14
Tefzel tubing, 0.03 ID x 0.062 OD
15
Holding Coil
6202918overview.fm
Table 1 Replacement Tubing Lengths for
Length
7.3 in. (18.5 cm)
24 in. (60 cm)
57 in. (145 cm)
61 in. (152.5 cm)
34.5 in. (87.5 cm)
57 in. (145 cm)
60 in. (152.5 cm)
8.3 in. (21 cm)
27.5 in. (70 cm)
7 in. (17.5 cm)
6 in. (16 cm)
26 in. (66 cm)
8.3 in. (21 cm)
Figure 1
From...
To...
Port 1
Colorimeter in
Port 2
Sample 1
Port 5
Standard 1
Port 6
Standard 2
Port 7
Waste
Port 8
Reagent 1
Port 9
Reagent 2
Port 10
Mixer module in
Mixer module out
Mixer drain
Two-way NC
Wash Solution
Two-way out
Autoburette bottom
Back pressure out
Colorimeter drain
Center port
Autoburette top
General Description
Section 1
Page 9

Advertisement

Table of Contents
loading

Table of Contents