  • Page 1 878 Instructions for Use...
  • Page 2: Table Of Contents

    English English CONTENTS 1. Device description ......3 5. Operation ....... . 11 1.1 Intended use .
  • Page 3: Device Description

    1.2 Description of function On the electronic personal scale, weight is detected by four load seca 878 cells. function can be used to determine the weight of babies and 2 in 1 toddlers. The child is held in an adult's arms during weighing to do so.
  • Page 4: Basic Safety Instructions

    You can find service partners in your area at www.seca.com or by sending an e-mail to service@seca.com. ► Only use original seca accessories and spare parts, otherwise seca will not grant any warranty. CAUTION! Patient hazard, malfunction ►...
  • Page 5: Preventing Injuries And Infections

    English Preventing injuries and infections WARNING! Injury from falls ► Ensure that the device is positioned firmly and level. ► Route connecting cables (if present) in such a way that neither user nor patient can trip over them. ► The device is not designed as a standing aid. Assist people with limited motor skills when they are getting up, e.g.
  • Page 6: Using The Measured Results

    NOTICE! Inconsistent measuring results ► Before you electronically save measurement values determined using this device and use them further (e.g. in seca PC software or in a hospital information system), make sure that the measurement values are plausible. ► If measurement values are transmitted to seca PC software or a...
  • Page 7: Overview

    (incl. rechargeable batteries). This prevents acid from leak- ing into the device. ► If acid leaked into the device, discontinue use. Have the device checked by an authorised seca service partner and repaired if necessary. 3. OVERVIEW 3.1 Controls...
  • Page 8: Symbols In The Display

    3.3 Marking on device and on the rating plate Text/symbol Meaning Model number Approval Type Type designation of design approval ( only) seca 878 Serial number, consecutive ProdID Product identification number, consecutive Follow instructions for use Electrical medical device, type B Insulated device, protection class II Value in mass units (verified models) •...
  • Page 9: Identification On The Packaging

    English 3.4 Identification on the packaging Protect from moisture Arrows indicate top of product. Transport and store in an upright position. Fragile Do not throw or drop. Permitted min. and max. temperature for transport and storage Permitted min. and max. moisture for transport and storage Open packaging here Packaging material can be disposed of through recycling...
  • Page 10: Establishing Power Supply

    The scale may overheat, catch fire, melt or short- circuit. ► Use only original seca plug-in power supply units with controlled 12 V output voltage. 1. Insert the plug required for your power supply into the power supply unit.
  • Page 11: Operation

    English 5. OPERATION 5.1 Weighing Starting the weighing procedure 1. Ensure that there is no load on the scale. 2. Press the key. start se[A appears in the display, then all the elements of the display are 0. 0 0 shown briefly.
  • Page 12: Switching Off The Scale

    4. To deactivate the HOLD function, press the hold 6. 5 0 symbol and the "HOLD" message are no longer displayed. Switching off the scale ► Press the key. start start key NOTE: The scale switches off automatically after a short time if there is no load on it.
  • Page 13: Sterilizing

    “What do I do if ...” in this document, you may not use the device. ► Have the device repaired by seca service or by an authorized service partner. ►...
  • Page 14: What Do I Do If

    Please only have servicing and repairs performed by an authorized service partner. ► You can find service partners in your area at www.seca.com or by sending an e-mail to service@seca.com. Have an authorized person perform reverification according to national legal regulations.
  • Page 15: Technical Data

    • electrical medical device, type B: Type of protection IP20 Operating mode Continuous duty 10.2 Weighing technology data Weighing technology data for seca 878 Verification in accordance with Directive 2014/31/EU Class III Maximum load 200 kg Minimum load 2 kg Accuracy on initial verification •...
  • Page 16: Optional Accessories

    We ask customers based abroad to contact their local sales agent directly in the case of warranty claims. 14.DECLARATION OF CONFORMITY seca gmbh & co. kg hereby declares that the product meets the terms of the applicable European directives. The unabridged declaration of conformity can be found at: www.seca.com.