C V P E R F O R M A N C E T E S T R E C O R; C C P E R F O R M A N C E T E S T R E C O R; Cv Load Effect (Load Regulation) - Agilent Technologies E3633A Service Manual

Hide thumbs Also See for E3633A:
Table of Contents

Advertisement

Chapter 3 Calibration Procedures
Performance Test Record for Agilent E3633A and E3634A
Performance Test Record for Agilent E3633A and E3634A
CV Performance Test Record
Test Description
CV Programming Accuracy @ 0
volts (DVM reading)
CV Readback Accuracy @ 0 volts
CV Programming Accuracy
@ Full Scale (DVM reading)
CV Readback Accuracy @ Full Scale (E3633A)

CV Load Effect (Load Regulation)

CV Source Effect (Line Regulation)
CV PARD (Normal mode)
Load Transient Response Time
CC Performance Test Record
Test Description
CC Programming Accuracy
@ 0 amps (I
)
O
CC Readback Accuracy @ 0 amps
CC Programming Accuracy
@ Full Scale (I
)
O
CC Readback Accuracy @ Full Scale (E3633A)
CC Load Effect (Load Regulation)
CC Source Effect (Line Regulation)
CC PARD (Normal mode)
CC PARD (Common mode)
*For Agilent E3633A Model
60
Models
Actual Result
both
both
(E3633A)
(E3634A)
(E3634A)
both
Models
Actual Result
both
both
(E3633A)
(E3634A)
(E3634A)
Maximum change: < (2.25 mA)* or (0.95 mA)**
Maximum change: < (2.25 mA)* or (0.95 mA)**
both
both
**For Agilent E3634A Model
Specifications
Upper Limit
Lower Limit
+0.0100 V
-0.0100 V
DVM + 0.0050 V
DVM - 0.0050 V
+20.0200 V
19.9800 V
+50.0350 V
49.9650 V
DVM + 0.0150 V
DVM - 0.0150 V
DVM + 0.0300 V
DVM - 0.0300 V
Maximum change: < (4 mV)* or (7 mV)**
Maximum change: < (4 mV)* or (7 mV)**
< 3 mV p-p, (350 μV)* or (500 μV)** rms
< 50 μsec
Specifications
Upper Limit
Lower Limit
+0.0100 A
-0.0100 A
I
+ 0.0050 A
I
O
O
20.0500 A
19.9500 A
7.0230 A
I
+ 0.0350 A
I
O
O
I
+ 0.0155 A
I
O
O
< 2 mA rms
< 1.5 μA rms
- 0.0050 A
6.9770 A
- 0.0350 A
- 0.0155 A

Advertisement

Table of Contents
loading

This manual is also suitable for:

E3634a

Table of Contents