Press ▲▼ key of channel 2 or 4 to modify the pulse width between 50 µs and 300 µs.
4 seconds after pressing the last key, the device automatically returns to the stimu-
lation mode. This is indicated by an acoustic signal.
The stimulation can then be started by increasing the intensity with the ▲▼ keys
of each connected channel.

Factory settings

Press the key for 4 seconds to retrieve the factory settings and delete the chan-
ged parameters in the AS SUPER 4 digital. This is indicated by an acoustic signal.

Technical Data

Technical specifications

Four-channel stimulator with electrically insulated channels, constant current cha-
racteristic, output short circuit control element (AKS), 37 integrated programs (7
of which are subject to modification of frequency, pulse width, therapy duration).
Output current :
Frequency range:
Impulse width:
Power supply:
Dimensions:
Weight:
There may be up to +/- 5% tolerance for all specification.

Description of the symbols

Attention: Read accompanying documents, especially user manual!
Type BF applied part
Date of manufacture
Manufacturer
Catalogue number
Serial number
max 20 mA (with a Load resistor of 1kΩ)
1-120 Hz
50-300 µs
mains supply transformer or 4 x 1.5 V battery (AAA)
17.5 x 11.3 x 4.5 cm
approx. 191 g (without batteries)