Troubleshooting; Technical Specifications - MEDIX AC2000 Instructions Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
en

8. Troubleshooting

Problem
Air compressor will not
operate when switched on.
Nebuliser chamber bubbles
or produces little or no mist.
Time taken to nebulise
significantly increases.

9. Technical Specifications

Country
Operating Voltage
Power Consumption
Fuse
Normal Flow
Weight
Dimensions
Type Rating
Sound Pressure Level 60 dbA approx.
*1 kPa = 0.01bar or 0.145psi
8
Action
If the green light in the mains switch is on, yet the motor does not
function, there is an internal fault. The machine should be returned
to the point of purchase or an authorised person for examination.
If the green light(s) on the switches are off:
1. Check mains lead is securely connected each end.
2. For machines with a 3 pin plug (UK) remove and check the fuse
in the plug. Fuse - 3.15AL 25mm long.
3. Check mains fuse in the control panel. Fuse - T1AL 20mm long.
4. Disconnect unit from electricity supply. Check mains fuse in
the mains inlet socket.
1. Check inlet filter. Replace if necessary.
2. Wash nebuliser chamber. Replace if problem still occurs.
3. Check driveline tubing is not kinked or leaking. Replace if
necessary.
4. Service compressor.
AC2000
Econoneb
UK/EU
UK/EU
220V-240V / 50Hz
220V-240V / 50Hz
62VA (260mA)
64VA (270mA)
1 x T1AL 20mm long
1 x T1AL 20mm long
min flow 7 L/min, max
min flow 7 L/min, max
flow 9 L/min @ 138 kPa* flow 9 L/min @ 138 kPa* flow 10.5 L/min @ 138 kPa*
2.8kg (w/o starter pack) 2.7kg
375 x 235 x 115mm
210 x 185 x 185mm
Class II BF Continuous
Class II BF Continuous
63 dbA approx.
Turboneb 2
UK/EU
230V / 50Hz
68VA (296mA)
1 x T1AL 20mm long
min flow 9.5 L/min, max
2.7kg
210 x 185 x 185mm
Class II BF Continuous
63 dbA approx.