Advertisement

5. When the Calibration/Point Set screen opens,
OK is highlighted — press Enter again.
6. When the Fluid Level - Calibration screen
re-opens, select Full from the Fluid level drop-
down, then select Calibrate and press Enter.
7. When the Calibration/Point Set screen opens,
OK is highlighted — press Enter again.
8. The calibration for the fluid level is now
complete — press Menu twice to navigate
back to the Device List screen.