EPS Bio Technology EasyPlus mini R13N User Manual page 39

Self-monitoring blood glucose system
Hide thumbs Also See for EasyPlus mini R13N:
Table of Contents

Advertisement

The test result
is among 20~60
mg/dL.
Action:
Please
contact your
healthcare
professional.
The test result is
lower
than
20
mg/dL.
Test again.
Action:
If the result is still
the same, please
contact your
healthcare
professional.
The "Ht" icon appears.
Temperature is too high,
outside the required
range of 50°F ~104°F
(10°C ~40°C). This alerts
users that an incorrect
result may occur if the
test continues.
Relocate the meter
Action:
to a location with
temperature between 50°F
~104°F (10°C ~40°C).
39
The "Lt" icon appears.
Temperature is too low,
outside the required
range of 50°F ~104°F
(10°C ~40°C). This alerts
users that an incorrect
result may occur if the
test continues.
Relocate the meter
Action:
to a location with
temperature between 50°F
~104°F (10°C ~40°C).