Troubleshooting Checks
Was the test strip exposing to open air
for a long period of time?
Was the test strip bottle capped tightly? This will humidify strips inside. Replace
Was the meter functioning well?
Is the control solution expired or
contaminated?
Were test strips and control solutions
stored in cool, dry places?
Did you follow the testing steps
properly?
Repeat the control test with properly stored
strips.
the test strips.
You can use the check strip to verify the
meter's functions.
Replace with new control solution to check
the performance of SMBG
Repeat the control test with properly
stored strips or control solutions.
Read Chapter 2 "Control Solution Testing"
and test again. If you still have problems,
please contact your local dealer for repair.