Selecting the Output Function
SELECTING THE OUTPUT FUNCTION
The C H A and CH
B
Keys
The CH A and CH B keys select the output waveform for the channel A and channel
B outputs, respectively. Each time the CH A or CH B key is pressed, the output wave-
form sequences through the OFF,
-
(sine wave),
%
(square wave), and DC outputs.
The selected output is indicated by an illuminated indicator. If the pulse mode is selected,
the sine wave and DC only output is removed from the output waveform selection se-
quence. The blue high voltage (HV) legends printed below these keys are active only
if the high voltage option (Option 002) is installed.
The Off, - ,
TI,
and DC Indicators
When all function indicators are extinguished, the output channel is disabled.
HP-16 Examples:
"
FCNAO"
or
"FCNA OFF"
"FCNBO"
or
"FCNB OFF"
The
-
(sine wave) indicator illuminates when a sine wave output is enabled. If the pulse
mode is selected, the sine wave output is removed from the output waveform selection
sequence.
HP-I6 Examples:
"FCNA1"
or
"FCNA SIN"
"FCNB 1 "
or
"FCNB SIN"