HP 3326A Operating And Reference Manual page 27

Two-channel synthesizer
Table of Contents

Advertisement

Data Entry and Modification
- -
-
*
-
DSHV
115
(Discrete Sweep High Voltage) The high voltage option is
enabled and a discrete frequency element frequency ex-
ceeds
1
MHz.
DSML
116
(Discrete Sweep ModuLation) Tee chanqel B frequency
exceeds ihe 5 kHz rnternal Pm limit or exceeds tne 100
kHz internal AV lrmit dcirir,g a discrete frequency sweep.
DSMD
117
(Discrete Sweep MoDe) The discrete frequency sweep ele-
ments in memory are incom~atible with the mode selected.
P OF
123
(Phase OFfset) Cannot cfear phase offset of channel A.
130, 136, 138
(High Voltage) The high voltage (HV) option cannot be ena-
bled because the output frequency 1 s qreater than 1 MHz,
CRPT
160
(CoRruPT) An error is detected in an instrument state
recalled from memory. The instrument state is replaced
(B OverLoad) The channel B output
IS
overloaded The load
impedance for the channel B output is too low or a volt-
AVCO
173
(channel A Voltage Controlled Oscillator) The channel A
voltaqe controlled oscrllator is unlocked
sensed, but the HP 3326A cannot lock to it
MCAL
192
(Mod~lation CAL~Dration) An internal AM or PM calibration
is ~nsuccessful.
PCAL
191
(Phase CALibration) A phase calibration is unsuccessful
ACA
L
192
(Amplitude CALibration) An amplitude call~ration is un-
successfu!.
OCAL
193-1 94
(Offset CALibration) A dc offset calibration is unsuccessful
PASS
-
A self test is successfvl.
FA1 L
A self test is unsuccessful.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents