Sanyo VPC-CG11W Service Manual page 18

Dual camera
Table of Contents

Advertisement

3-6. The adjustment item which in necessary in part exchange
COMPL PWB CP1
COMPL PWB VF1
COMPL PWB ST1
COMPL PWB TB1
ASSY FPC, CA1
: Be sure to carry out the necessary adjustments after replacing the unit.
: Adjustment is possible from the menu setting screen of the camera and by using the calibration software.
3-7. Updating the firmware
Check the firmware version immediately after the CP1 board has been replaced. If an old version is being used, interfer-
ence and errors in operation may also occur. If an old version is being used, update it with a newer version.
3-13. Firmware uploading procedure. (Page 21)
Refer to
3-8. Adjust Specifications
1. Lens Adjustment (Infinity)
Camera
Preparation:
POWER switch: ON
If using a ready-made collimator, set to infinity.
Note:
Do not vibrate during the adjustment.
If readjusting after it has already been adjusted, wait for 15
minutes or more for the unit to cool down first.
Adjustment method:
1. Set a distance of 0.5-1.0 cm between camera lens and
collimator lens. Do not touch the each lens.
Lens
Lens
AWB
Adjust-
Adjust-
Adjust-
ment
ment
ment
(Infinity)
(1 m)
Collimator
CCD Black
CCD White
Point And
Point
White Point
Defect
Defect
Detect
Detect
Adjustment
Adjustment
In Lighted
2. Set the camera so that it becomes center of the screen in
the collimator (zoom wide and tele).
3. Double-click on the DscCalDi.exe.
4. Select "Infinity Cal." on the LCD "Test", and click the "Yes".
5. Lens infinity adjustment value will appear on the screen.
6. Click the OK.
2. Lens Adjustment (1m) is carried out after this adjust-
ment.
Dsc Calibration
Infinity calibration :
AF_TEMP_AD_I: 413
PZ_BR: 4
AF_BR: 4
AF_I_WIDE: 4
AF_I_MID1: 4
AF_I_MID2: 13
AF_I_TELE: 39
IRIS_GAIN: 100
IRIS_OFFSET: 31
-
-
18