4 Using the control module
Viscosity Delay
Sets how many seconds the plunger pauses at the top of the pump and
injection strokes.
Depth offset
Sets the depth offset of the syringe needle.

4.3 Method Setup - GC

4.3.1 Oven/Zones

This tab allows to set the parameters for oven and temperature zones.
Particular fields may vary, according to the configuration.
Oven : Max
Sets the maximal allowed temperature for the column oven.
Note:
Please note that the maximum value may differ for certain GC models.
Equilibration
Sets the time of oven temperature equilibration.
Gradient Table
Sets the Heat Rate, desired temperature (Final Temp) and time to hold
this temperature (Hold Time). The Total Time column will be calculated
automatically. In other words, for isothermal part of the temperature
gradient, the temperature set in Final Temp will be kept for the interval set
Fig 7: Method Setup - GC - Oven/Zones
- 11 -