Using The Control Module; Method Setup - As - DataApex M070C Manual

Control module
Table of Contents

Advertisement

Spark Marathon

4 Using the control module

New

Method Setup - AS

setting of the Spark Marathon autosampler control method.
4.1 Method Setup - AS
The Method Setup - AS tab defines the properties of the Spark Marathon
autosampler instrument method. The fields available in the dialog may vary
according to the autosampler model connected and installed options.
Analysis Time [min.]
Sets the time after which the autosampler will start to inject next sample.
Note:
As each injection is sent from Clarity Sequence Table by separate command,
the Analysis Time field is not used and injections are controlled by Clarity when
the sampler is operated by Clarity. The field is nonetheless active to provide
compatibility with eventual manual control of the AS.
Loop Volume [μl]
Defines the value of the loop volume used for the injection (in μl). If the installed
loop volume is different, the injection reproducibility may be impaired. Possible
values range between 0 and 999 μl.
Auxiliaries
Sets the operation with two external contacts of the Spark Marathon autosampler.
Each auxiliary external contact has its own row, on which can be once switched on
and once switched off based on the time set in minutes. Available values for the
tab appears in the
Fig. 6: Method Setup - AS
- 7 -
4 Using the control module
Method Setup
dialog, enabling the

Advertisement