Method Setup - AS

the setting of the Spark Marathon autosampler control method.
4.1 Method Setup - AS
The Method Setup - AS tab defines the properties of the Spark Marathon
autosampler instrument method. The fields available in the dialog may vary
according to the autosampler model connected and installed options.
Analysis Time [min.]
Sets the time after which the autosampler will start to inject next sample.
Note:
As each injection is sent from Clarity Sequence Table by separate
command, the Analysis Time field is not used and injections are
controlled by Clarity when the sampler is operated by Clarity. The field
is nonetheless active to provide compatibility with eventual manual
control of the AS.
Loop Volume [μl]
Defines the value of the loop volume used for the injection (in μl). If the
installed loop volume is different, the injection reproducibility may be
impaired. Possible values range between 0 and 999 μl.
Auxiliaries
Sets the operation with two external contacts of the Spark Marathon
autosampler. Each auxiliary external contact has its own row, on which can
tab appears in the
Fig 6: Method Setup - AS
- 7 -
4 Using the control module
Method Setup
dialog, enabling