Specifications

Power supply
Display mode
Measurement mode
Measurement range
Accuracy
Normal working condition
Storage & transportation condition
Measurement perimeter of the wrist
Net Weight
External dimensions
Attachment
Mode of operation
Degree of protection
Protection against ingress of water
Software version
Device classification
WARNING: No modification of this equipment is allowed
3.7V 420mAH Built-in rechargeable lithium-ion battery, 5V / 1A USB
Digital LCD V.A.46.5x36.5mm
Oscillographic testing mode
Rated cuff pressure:
0kpa-40kpa (0mmHg-300mmHg) Measurement pressure: 5,3 kPa-
30,7kPa (40mmHg-230mmHg)
pulse value:(40-199)beat/minute
Pressure:
5° C – 40° C within±0.4kpa(3mmHg)
pulse value:±5%
Temperature: 5° C to 40° C Relative humidity ≤85%
Atmospheric pressure: 86kPa to 106kPa
Temperature:-20° C to 60° C RH: 10% to 93% Atmospheric pressure: 50kPa to
106kPa
About 13.5cm-21.5cm
Approx.110g
Approx.79.8×72.5×13.2mm
USB cable, user manual
Continuous operation
Type BF applied part
IP22, It means the device could protected against solid foreign objects of
12.5mm and greater, and against vertically falling water drops when
ENCLOSURE tilted up to 15°
V01
Battery Powered Mode:
Internally Powered ME Equipment
AC Adaptor charged Mode: Class II ME Equipment