Retrieving Logged Data (1100D); G. Troubleshooting & Maintenance - Digitron 1100T Instruction Manual

Table of Contents

Advertisement

 
 
 
 
 
 
F. Retrieving Logged Data (1100D Only)
1.
Install the 1100D PC interface software on the computer.
2.
Attach the 1100D interface dock to the 1100D and connect the
USB cable from the dock to the PC.
3.
The 1100D interface software will automatically detect the 1100D
and begin communication.
4.
Through the software all data logged can be retrieved.
 
 
 
 
 
 
 
34
 
G. Troubleshooting and Maintenance
 
Troubleshooting
Problem
Probable Causes
 
Does not power up
• Dead or defective battery
• Broken wire from battery
snap to PCB
 
Temperature reads
• Very weak battery that will
high
not turn on the low battery
indicator on the LCD
 
Only reads OPEN
• Temperature probe not connected
• Temperature probe defective
Maintenance
1. Battery Replacement: The 1100D will display a
battery symbol when the internal 1.5 Volt batteries
needs replacement.
a. Disconnect the temperature probe.
b. Loosen the 4 battery cover screws located on the
back cover.
c. Remove the battery cover.
d. Remove old batteries and replace with new batteries.
Observe the correct polarity.
e. Reassemble the instrument in reverse
order from above.
2. Cleaning your 1100T/D:
Use a mild detergent and slightly damp cloth to clean the
surfaces of the unit.
35

Advertisement

Table of Contents
loading

This manual is also suitable for:

1100d

Table of Contents