RATIO
3.13
The
principle of
operation
is
very
similar to
frequency measurement, except
that
the time base output
is
not
derived from the frequency standard.
Instead,
the
lower
of the
frequencies
to
be compared
is
applied
to
the
'B'
Input
and
divided
in
the time
base.
The
higher frequency
is
applied
to
the
'A'
Input
and
fed via the
main
gate
to the
counter.
A
waveform
from
the
time base
decades
(derived from the
'B'
Input)
is
selected
to
gate
the
'A'
channel
signal
by
depressing
an
appropriate
Range
('n')
push-button.
Since
the
time base output
is
a divided version of the
'B'
signal,
the readout gives
a count
which
corresponds
to the
ratio
of the
two
input frequencies, multiplied
by
the factor
'n'
.
The
particular
'n'
factor
used
is
that
shown
adjacent
to
the selected push-button.
SHAPER
Ratio
Measurement
Fig.
3.
5